Applications
The PACT Application
268
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
Selecting/Deselecting Power Meters
8
Move to the Module Selection box. As a default, all Power Meters are selected,
this means that all Power Meters will log power measurements. Press [Edit].
9
To deselect a Power Meter channel, move to the Power Meter as denoted by
slot and channel numbers, see “
Slot and Channel Numbers
” on page 41, press
[Unset].
10
To select a Power Meter channel, press [Edit], move to the Power Meter as
denoted by slot and channel numbers, see “
Slot and Channel Numbers
” on
page 41, press [Set].
11
Perform steps until you have selected the required Power Meters for your
application and press Enter.
Setting Sweep Parameters
12
Move to [
λ
Start], press Enter, enter the start wavelength of the sweep, and
press Enter.
13
Move to [
λ
Stop], press Enter, enter the stop wavelength of the sweep, and
press Enter.
14
Move to [Step], press Enter, enter the step size for the sweep, and press Enter.
15
Move to [P max], press Enter, enter the maximum power in dBm for the sweep,
and press Enter. The instrument chooses a constant power level for the
measurement using the following constraints:
•
the power will be less than or equal to the [P max] value you set and
•
the power will be less than or equal to the maximum power possible for the
chosen Tunable Laser module across the chosen wavelength range.
If no Tunable Laser module is selected, you cannot exit by pressing Enter.
N O T E
If no Power Meter channel is selected, you cannot exit by pressing Enter.
N O T E
The step size defines:
• the size of the change in the wavelength for each step of a stepped sweep and
• the wavelength interval between power measurements, the first power meas-
urement is taken at [
λ
Start] and the last must be taken at [
λ
Stop].
N O T E
Содержание 8163B
Страница 4: ......
Страница 20: ...20 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 22: ...22 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 62: ...Getting Started A Sample Session 62 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 82: ...Additional Information How to Connect a Printer 82 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 108: ...Power Measurement How to Measure Power 108 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 118: ...Laser Sources How to Use Laser Source Modules 118 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 158: ...Tunable Lasers How to Use Auxiliary Functions 158 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 318: ...Installation and Maintenance Claims and Repackaging 318 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 334: ...Accessories GPIB Cables and Adapters 334 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 368: ...Performance Tests Test Record 368 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 378: ...Performance Tests Test Record 378 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 450: ...Cleaning Instructions Other Cleaning Hints 450 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 462: ...Firmware Upgrades Firmware Upgrade Process 462 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 465: ......