Return Loss Measurement
Return Loss Measurement
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
177
Calibration using the Agilent 81000BR
Reference Reflector
You can use the agilent 81000BR Reference Reflector to make a
Reflectance Calibration, and a Termination Calibration.
Reflectance Calibration
Use this procedure to calibrate the Return Loss module against a
component of known return loss.
1
Making sure all the connectors are clean, set up the instrument as shown in
Figure 93 if you are using an External Source.
Figure 93 Reflectance Calibration - External Source
or Figure 94 if you are using an internal source
(Agilent 81611A/2A/3A/4A Return Loss modules only).
Figure 94 Reflection Calibration - Internal Source
Tip: For best results and higher repeatability, fix the cable.
2
Attach a component with a known return loss to the end of the patchcord.
Tip: The Agilent 81000BR Back Reflector is such a component, offering a
return loss of 0.18
±
0.1 dB.
3
Move to the Return Loss channel, then press [Details]
4
Move to [FPDelta], press [Edit], set the value to 0.000, then press [OK].
Laser
Agilent 81113PC
Agilent 81000BR
Agilent 81000FI
Agilent 81000UM
8161x
8163B Lightwave Multimeter
Laser
Source
Agilent 81113PC
8161x
8163B Lightwave Multimeter
Agilent 81113PC
Agilent 81000BR
Agilent 81000FI
Agilent 81000UM
8161x
8163B Lightwave Multimeter
Содержание 8163B
Страница 4: ......
Страница 20: ...20 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 22: ...22 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 62: ...Getting Started A Sample Session 62 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 82: ...Additional Information How to Connect a Printer 82 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 108: ...Power Measurement How to Measure Power 108 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 118: ...Laser Sources How to Use Laser Source Modules 118 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 158: ...Tunable Lasers How to Use Auxiliary Functions 158 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 318: ...Installation and Maintenance Claims and Repackaging 318 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 334: ...Accessories GPIB Cables and Adapters 334 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 368: ...Performance Tests Test Record 368 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 378: ...Performance Tests Test Record 378 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 450: ...Cleaning Instructions Other Cleaning Hints 450 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 462: ...Firmware Upgrades Firmware Upgrade Process 462 Agilent 8163A B 8164A B 8166A B Mainframes Sixth Edition ...
Страница 465: ......