72
708-DS / 709-DS Operator’s Manual
4
Operating the 708-DS / 709-DS
Method Properties Screen 3
Table 4
Method Properties Screen 2
Option
Function
Enable Manifold
Enabling this option activates the optionally installed sampling
manifold. The manifold lowers the sampling probes into the solution,
allowing automated sample collection as well as the temperature probes
for automated temperature measurement.
Initial Temp
Enables recording and printing of the initial test temperature.
Final Temp
Enables recording and printing of the final test temperature.
Manifold Down
Lead Time
This is the amount of time that the manifold lowers prior to the sample
time point. This allows for equilibration of the temperature probes
before vessel temperature measurement and for priming the system
when an external pump is connected.
Manifold Down
Duration
This is the amount of time that the manifold remains at its sampling
position. The sum of the Lead Time and Duration must be greater than
or equal to 30 seconds for accurate temperature readings.
Содержание 708-DS
Страница 1: ...Agilent Technologies 708 DS 709 DS Dissolution Apparatus Operator s Manual ...
Страница 10: ...10 708 DS 709 DS Operator s Manual Tables This page was intentionally left blank except for this message ...
Страница 125: ...Maintenance and Troubleshooting 5 708 DS 709 DS Operator s Manual 125 Figure 29 Resistance Value of Temperature Probe ...
Страница 133: ...Maintenance and Troubleshooting 5 708 DS 709 DS Operator s Manual 133 Figure 37 Feeding Paper ...
Страница 136: ...136 708 DS 709 DS Operator s Manual 5 Maintenance and Troubleshooting ...
Страница 140: ...138 708 DS 709 DS Operator s Manual Index This page was intentionally left blank except for this message ...