![Agilent Technologies 708-DS Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/agilent-technologies/708-ds/708-ds_operators-manual_2867627043.webp)
Setting Up the 708-DS / 709-DS
3
708-DS / 709-DS Operator’s Manual
43
Figure 21
Accessory Installation
Setting Basket (Apparatus 1) Heights
1
Ensure the drive unit is fully raised and the basket shafts are pushed
up sufficiently.
2
Clip the basket height gauge provided onto the bottom of the basket
shaft.
3
Lower the drive unit to its operating position (until it stops).
4
With the shaft locking collar loosened, carefully lower the shaft until
the bottom of the basket height gauge rests against the bottom of the
vessel.
5
Ensure the shaft locking collar is flush against the top of the spindle
assembly by rotating each shaft until resistance is met.
6
Tighten the shaft locking collar securely.