708-DS / 709-DS Operator’s Manual
5
Maintenance and Troubleshooting
Water Bath Temperature Probe Accuracy Test
The temperature probe can be tested using a high-quality ohmmeter
capable of at least four-digit resolution and a known temperature bath.
Both the multi-meter and the bath temperature must be traceable to a
known reference standard such as NIST.
1
Place the probe into a water bath with a known temperature and
allow several minutes for the probe to equilibrate.
2
The resistance is measured by attaching the leads of the multi-meter
to the top and shank of the ¼-inch phone plug.
3
Refer to the table on Page
to find the resistance value of the
probe at the bath temperature. Temperatures between the values
listed may be interpolated.
4
Probes are interchangeable and manufactured with a tolerance of +/-
0.1 °C. Probes found to be outside of the tolerance of +/- 0.1 °C
should be replaced.
Содержание 708-DS
Страница 1: ...Agilent Technologies 708 DS 709 DS Dissolution Apparatus Operator s Manual ...
Страница 10: ...10 708 DS 709 DS Operator s Manual Tables This page was intentionally left blank except for this message ...
Страница 125: ...Maintenance and Troubleshooting 5 708 DS 709 DS Operator s Manual 125 Figure 29 Resistance Value of Temperature Probe ...
Страница 133: ...Maintenance and Troubleshooting 5 708 DS 709 DS Operator s Manual 133 Figure 37 Feeding Paper ...
Страница 136: ...136 708 DS 709 DS Operator s Manual 5 Maintenance and Troubleshooting ...
Страница 140: ...138 708 DS 709 DS Operator s Manual Index This page was intentionally left blank except for this message ...