Vessel Installation
1
Press
Drive
to move the drive unit to the fully raised position.
2
Carefully install the dissolution vessels in the vessel plate.
3
For the 708-DS vessel, engage the clips for each position over the
rim of the vessel to secure it in place.
For the 709-DS vessel, open the clip, install the vessel, and close the
clip.
4
If necessary, place the TruAlign blank vessel position covers on the
vessel plate openings at Position 7 and Position 8. Rotate the two
clips for each position over the cover to secure it in place.
NOTE
Bubbles on the surface of the vessel, like the one centered in the figure
below, are not a manufacturing defect and will not affect performance.
Figure 11
Vessel Bubble
CAUTION
Ensure the connection pins in the vessel socket are aligned with the
holes on the vessel.
Содержание 708-DS
Страница 1: ...Agilent Technologies 708 DS 709 DS Dissolution Apparatus Operator s Manual ...
Страница 10: ...10 708 DS 709 DS Operator s Manual Tables This page was intentionally left blank except for this message ...
Страница 125: ...Maintenance and Troubleshooting 5 708 DS 709 DS Operator s Manual 125 Figure 29 Resistance Value of Temperature Probe ...
Страница 133: ...Maintenance and Troubleshooting 5 708 DS 709 DS Operator s Manual 133 Figure 37 Feeding Paper ...
Страница 136: ...136 708 DS 709 DS Operator s Manual 5 Maintenance and Troubleshooting ...
Страница 140: ...138 708 DS 709 DS Operator s Manual Index This page was intentionally left blank except for this message ...