![Agilent Technologies 708-DS Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/agilent-technologies/708-ds/708-ds_operators-manual_2867627035.webp)
Setting Up the 708-DS / 709-DS
3
708-DS / 709-DS Operator’s Manual
35
Sample Manifold Installation
1
From the Main screen, press
Menu
>
Diagnostics
.
2
Position the sample manifold so the center screw is aligned with the
center receptor beneath the drive unit. Ensure the left and right
alignment shafts are also positioned correctly.
Figure 12
Positioning the Manifold for Installation
3
Push the manifold up into the drive unit until resistance is met.
4
While gently lifting upward, press
Load
from within the Manifold
Control box. The internal drive should slowly lift the manifold to its
fully raised position.
5
If the unit is equipped with AutoTemp, attach the two 9-pin RS232
cables to the underside of the drive unit.