30
708-DS / 709-DS Operator’s Manual
3
Setting Up the 708-DS / 709-DS
EVT Probe Installation (708-DS)
1
If applicable, insert the EVT Probe cable into the VESSEL TEMP
jack on the rear of the dissolution apparatus drive unit.
2
Place the vessel temperature probe in the hole in the vessel plate, at
the left side rear of the apparatus.
Power Cord Connections
1
Connect or verify connections of power cords to all components.
2
Plug the cords into electrical outlets of the appropriate voltage.
708-DS / 709-DS Initial Power Up
1
Turn on the power of the dissolution apparatus using the switch
located on the left side of the drive unit.
2
Ensure the self test conducted upon initialization passes, which is
indicated by a green check mark.
CAUTION
Ensure the correct voltage has been supplied prior to connecting the
power cables.
CAUTION
Ensure the retaining ties have been removed. Failure to do so could
result in significant damage to the instrument.
Содержание 708-DS
Страница 1: ...Agilent Technologies 708 DS 709 DS Dissolution Apparatus Operator s Manual ...
Страница 10: ...10 708 DS 709 DS Operator s Manual Tables This page was intentionally left blank except for this message ...
Страница 125: ...Maintenance and Troubleshooting 5 708 DS 709 DS Operator s Manual 125 Figure 29 Resistance Value of Temperature Probe ...
Страница 133: ...Maintenance and Troubleshooting 5 708 DS 709 DS Operator s Manual 133 Figure 37 Feeding Paper ...
Страница 136: ...136 708 DS 709 DS Operator s Manual 5 Maintenance and Troubleshooting ...
Страница 140: ...138 708 DS 709 DS Operator s Manual Index This page was intentionally left blank except for this message ...