B - Performance, Calibration, and Configuration
160
To access the front panel commands for the configuration function, the calibration state must be enabled.
If a calibration password is set, the password must be provided to enable calibration. Once calibration is
enabled, the front panel will be in Meter mode. The configuration function can only be invoked while in
meter mode. If for some reason the front panel is not in meter mode, press the
Meter
key. To access the
configuration menu, you must then simultaneously press the
Local
and the
Recall
keys.
The Configuration function consists of the following menu commands.
Press this key to access the calibration menu.
Simultaneously press these keys to access the configuration menu.
Display
Command
Function
CFG EXIT
Exits the configuration menu and returns to meter mode. Also
exits calibration mode. Changes not previously saved are lost.
CFG:FACTORY
Returns the *RST state to the factory default settings.
CFG:COMP <char>
Sets the output compensation mode.
CFG:COUP <char>
Sets the output coupling mode.
CFG:RELAY1<char>
Sets the output 1 relay mode (Option 521 units only).
CFG:RELAY2<char>
Sets the output 2 relay mode (Option 521 units only).
CFG:OVP <char>
Sets the overvoltage protection mode.
CFG:SAVE
Stores the current settings in nonvolatile RAM.
CFG:UNDO
Cancels any changes that have been made but not yet saved.
Notes:
char = a character string parameter
Use and to scroll through the menu commands.
Use and to scroll through the menu parameters.
Once the configuration function is started, the front panel will remain in the configuration mode. No other
functions can be accessed from the keypad. To exit the configuration function, the CFG:EXIT command
must be executed. Each configuration command is only executed after the
Enter
key is pressed. An
“OKAY” message is displayed to indicate that an action has been taken. When the configuration
parameters are changed, they are not immediately stored into non-volatile memory. To store the change
into non-volatile memory, the CFG:SAVE command must be executed
Note that the unit is put into the present *RST state upon entering and exiting the configuration function.