![Agilent Technologies 66319B Скачать руководство пользователя страница 153](http://html1.mh-extra.com/html/agilent-technologies/66319b/66319b_user-manual_2868397153.webp)
Performance, Calibration, and Configuration - B
153
Model 66319B/D Output 2
Report No _______________ Date __________________
Test Description
Minimum
Specification
Results
Maximum
Specification
CONSTANT VOLTAGE TESTS
Voltage Programming and Readback
Output 2 Low Voltage @ 0 V
Output 2 Front Panel Display Readback
Output 2 High Voltage @ Full Scale
Output 2 Front Panel Display Readback
−
40 mV
Vout2
−
15 mV
11.936 V
Vout2
−
39 mV
__________
__________
__________
__________
+ 40 mV
Vout2
+ 15 mV
12.064 V
Vout2
+ 39 mV
Load Effect
−
1.6 mV
__________
+
1.6mV
Source Effect
−
0.4 mV
__________
+ 0.4 mV
PARD (Ripple and Noise)
Peak-to-Peak
RMS (with phone capacitance <6
µ
F)
N.A.
N.A.
__________
__________
+ 6 mV
+ 1 mV
Transient Response
1
Time in <400
µ
s
−
20 mV
__________
+ 20 mV
CONSTANT CURRENT TESTS
Current Programming and Readback
Output 2 Low Current @ 0 A
Output 2 High Current @ Full Scale
Output 2 Front Panel Display Readback
−
4.5 mA
1.4925 A
Iout2
−
6 mA
__________
__________
__________
+ 4.5 mA
1.5075 A
Iout2 +
6 mA
Current Sink (-0.032A @ 7.5V) Readback
Isink
−
25 mA
__________
Isink
+ 39 mA
PARD (Current Ripple and Noise)
RMS
N.A.
__________
+ 2.0 mA
Load Effect
−
0.375 mA
__________
+ 0.375 mA
Source Effect
−
0.25 mA
__________
+
0.25 mA
1
Following a 0.75A to 1.5A load change
Model 66319D/66321D DVM Input
Report No _______________ Date __________________
Test Description
Minimum
Specification
Results
Maximum
Specification
DVM VOLTAGE CALIBRATION VERIFICATION
Positive Voltage Measurement
Negative Voltage Measurement
−
15mV
−
6.8mV
_________
_________
+
15mV
+
6.8mV