B - Performance, Calibration, and Configuration
154
Performing the Calibration Procedure
NOTE:
The calibration procedure can only be performed from the front panel or using the SCPI
language commands.
Table B-1 lists the equipment required for calibration. Figure B-1 shows the test setup.
You do not have to do a complete calibration each time. If appropriate, you may calibrate only the
voltage or current and proceed to "Saving the Calibration Constants". However, the voltage or current
calibration sequence must be performed in its entirety. The following parameters may be calibrated:
♦
voltage programming and measurement
♦
current programming and measurement
♦
1 A and 0.02A range measurements
♦
ac current measurement
♦
resistance
programming
♦
internal
DVM
Front Panel Calibration Menu
The Entry keypad is used for calibration functions.
Press this key to access the calibration menu.
Display Command
Function
CAL ON <value>
Turns calibration mode on when the correct password
value is entered.
CAL OFF
Turns calibration mode off
CAL:LEV <char>
Advance to next step in sequence (P1 or P2).
CAL:DATA <value>
Enter an external calibration measurement.
CAL:VOLT
Begin voltage calibration sequence
CAL:VOLT2
Begin output 2 voltage calibration sequence
CAL:RES
Begin resistance calibration
CAL:VOLT:PROT
Not
required
CAL:CURR
Begin high range current calibration sequence
CAL:CURR2
Begin output 2 current calibration sequence
CAL:CURR:MEAS:LOW
Begin 0.02A range current measurement calibration
CAL:CURR:MEAS:R3
Begin 1A range current measurement calibration
CAL:CURR:MEAS:AC
Begin ac current calibration sequence
CAL:DVM
Begin DVM calibration sequence
CAL:SAVE
Saves the calibration constants in non-volatile memory.
CAL:PASS <value>
Set new calibration password.
DATE <char>
Displays the calibration date. (0 if none is supplied)
Notes:
value = a numeric value
char = a character string parameter
Use and to scroll through the menu commands.
Use and to scroll through the menu parameters.
Use and to select a digit in a numeric entry field.