Performance, Calibration, and Configuration - B
155
Front Panel Calibration Procedure
These procedures assume you understand how to operate front panel keys (see chapter 5). Make sure the
sense terminals are directly jumpered to the output terminals.
Enable Calibration Mode
Action
Display
1.
Reset the unit by selecting
Output
, scrolling to *RST and pressing
Enter
.
*RST
2.
Press
Output On/Off
to enable the output.
00.003V 0.0006A
3.
To begin calibration press
Shift Cal
, scroll to CAL ON and press
Enter
.
CAL ON 0.0
4.
Enter the calibration password from Entry keypad and press
Enter
. lf the
password is correct the Cal annunciator will come on.
If CAL DENIED appears, then an internal switch has been set to prevent the
calibration from being changed.
lf the password is incorrect, an error occurs. If the active password is lost, the
calibration function can be recovered by setting an internal switch that defeats
password protection (contact the factory for details).
CAL DENIED
OUT OF RANGE
Voltage Programming and Measurement Calibration
Action
Display
5.
Connect the external DMM (in voltage mode) directly to output 1.
Do not connect the load resistor or current shunt. Select output 1.
6.
Press
Shift Cal
, scroll to CAL VOLT, and press
Enter
.
CAL:VOLT
7.
Press
Shift Cal
, scroll to CAL LEV, and press
Enter
to select the first
calibration point.
CAL:LEV P1
8.
Press
Shift Cal
, scroll to CAL DATA, press
Enter Number
, and enter the
voltage value displayed on the DMM. Press
Enter
.
CAL:DATA 0.00
9.
Press
Shift Cal
, scroll to CAL LEV, use
Æ
to scroll to P2 (the second
calibration point), and press
Enter
.
CAL:LEV P2
10.
Press
Shift Cal
, scroll to CAL DATA, press
Enter Number
, and enter the
second voltage value displayed on the DMM. Press
Enter
.
CAL:DATA 0.00
Agilent 66319B/D Output 2 Voltage Programming and Measurement Calibration
Action
Display
11.
Connect the external DMM (in voltage mode) directly to output 2.
Do not connect the load resistor or current shunt. Select output 2.
12.
Press
Shift Cal
, scroll to CAL VOLT2, and press
Enter
.
CAL:VOLT2
13.
Press
Shift Cal
, scroll to CAL LEV, and press
Enter
to select the first
calibration point.
CAL:LEV P1
14.
Press
Shift Cal
, scroll to CAL DATA, press
Enter Number
, and enter the
voltage value displayed on the DMM. Press
Enter
.
CAL:DATA 0.00
15.
Press
Shift Cal
, scroll to CAL LEV, use
Æ
to scroll to P2 (the second
calibration point), and press
Enter
.
CAL:LEV P2
16.
Press
Shift Cal
, scroll to CAL DATA, press
Enter Number
, and enter the
second voltage value displayed on the DMM. Press
Enter
.
CAL:DATA 0.00