B - Performance, Calibration, and Configuration
158
Agilent 66321D, 66319D DVM Calibration
Action
Display
51.
Connect the DVM inputs directly to output 1. Connect the external DMM to
the DVM inputs as shown in figure B-1e. Do not connect the Agilent 3478
DMM.
52.
Press
Shift Cal
, scroll to CAL DVM, and press
Enter
.
CAL:DVM
53.
Press
Shift Cal
, scroll to CAL LEV, and press
Enter
to select the first
calibration point.
CAL:LEV P1
54.
Press
Shift Cal
, scroll to CAL DATA, press
Enter Number
, and enter
the voltage value displayed on the external DMM. Numbers entered from
the external DMM should have an accuracy and resolution of 0.1 mV or
better. Press
Enter
.
CAL:DATA 0.00
55.
Reverse the output 1 connections. Move the - lead to the + DVM connector
pin, and move the + lead to the - DVM connector pin.
56.
Press
Shift Cal
, scroll to CAL LEV, use
Æ
to scroll to P2 (the second
calibration point), and press
Enter
.
CAL:LEV P2
57.
Press
Shift Cal
, scroll to CAL DATA, press
Enter Number
, and enter
the second voltage value displayed on the external DMM. Numbers entered
from the external DMM should have an accuracy and resolution of 0.1 mV
or better.
Because this second value is negative, you must place a minus sign in front
of the value. Pressing the
Enter Number
key the second time enters the
minus sign. Then press
Enter
.
CAL:DATA
−
0.00
Saving the Calibration Constants
WARNING:
Storing calibration constants overwrites the existing ones in non-volatile memory. If you
are not sure you want to permanently store the new constants, omit this step. The dc
source calibration will then remain unchanged.
Action
Display
58.
Press
Shift Cal
, scroll to CAL SAVE, and press
Enter
.
CAL:SAVE
59. Press
Shift Cal
, select CAL OFF, and press
Enter
to exit Calibration mode.
*RST and *RCL will also set the calibration state to OFF.
CAL OFF
Calibration Error Messages
Errors that can occur during calibration are shown in the following table.
Table B-3. GPIB Calibration Error Messages
Error Meaning
401
CAL switch prevents calibration (contact the factory for details).
402
CAL password is incorrect
403
CAL not enabled
404
Computed readback cal constants are incorrect
405
Computed programming cal constants are incorrect
406
Incorrect sequence of calibration commands