![Agilent Technologies 4395A Скачать руководство пользователя страница 432](http://html.mh-extra.com/html/agilent-technologies/4395a/4395a_manual-change_2867896432.webp)
Calibration
for
Network
Measurement
Reection
Tracking,
E
RF
Reverse
Directivity
,
E
DR
Isolation,
E
XR
Source
Match,
E
SR
Load
Match,
E
LR
Transmission
Tracking,
E
TR
Reection
Tracking,
E
RR
The
87511A,
B
S-parameter
T
est
sets
can
measure
both
the
forward
and
reverse
characteristics
of
the
test
device
without
the
need
to
manually
remove
and
physically
reverse
it.
With
these
test
sets
,
the
full
two-port
error
model
illustrated
in
Figure
A
-40
eectively
removes
both
the
forward
and
reverse
error
terms
for
transmission
and
reection
measurements
.
The
87512A,
B
Transmission/Reection
T
est
kits
cannot
switch
between
forward
and
reverse
directions
,
so
the
reverse
error
terms
cannot
be
automatically
measured.
Therefore
,
with
the
one-path
two-port
calibration,
the
forward
error
terms
are
duplicated
and
used
for
both
forward
and
reverse
measurements
by
manually
reversing
the
test
device
.
Figure
A
-40.
Full
Two-P
ort
Error
Model
The
following
equations
show
the
full
two-port
error
model
equations
for
all
four
S-parameters
of
a
two-port
device
.
Note
that
the
mathematics
for
this
comprehensive
model
use
all
forward
and
reverse
error
terms
and
measured
values
.
Thus
,
to
perform
full
error
correction
for
any
one
parameter
,
all
four
S-parameters
must
be
measured.
A-56
Basic
Measurement
Theory
Содержание 4395A
Страница 8: ......
Страница 9: ......
Страница 10: ......
Страница 12: ......
Страница 13: ......
Страница 14: ......
Страница 16: ......
Страница 49: ...D 1 Manual Changes by Serial Number D 1 D 2 Manual Changes by Firmware Version D 1 Contents 23 ...
Страница 50: ......
Страница 58: ...Power Requirements Figure 2 1 Power Cable Supplied 2 6 Installation Guide ...
Страница 69: ...Connecting an Active Probe Figure 2 10 Connecting Test Fixture Installation Guide 2 17 ...
Страница 72: ......
Страница 76: ...Step 1 Preparing for the Measurement Figure 3 3 S Parameter Test Set Setup 3 4 Quick Start Guide ...
Страница 93: ...Step 2 Setting Up the 4395A Quick Start Guide 3 21 ...
Страница 130: ......
Страница 144: ......
Страница 154: ......
Страница 198: ...Calibration Required for the Impedance Analyzer Mode Figure 7 2 Connecting Test Fixture 7 16 Calibration ...
Страница 251: ...Measuring Zero Span Figure 8 28 Marker Time Analyzing the Measurement Results 8 49 ...
Страница 262: ......
Страница 278: ......
Страница 298: ...AM Signal Measurement SA where 1Mkr is the 1marker value shown in Figure 10 18 10 20 Examples of Applications ...
Страница 301: ...FM Signal Measurement SA Figure 10 20 Zooming Carrier Signal of FM Signal Examples of Applications 10 23 ...
Страница 374: ...Measurement accessories available 11855A 75 Type N Adapter Kit 11856A 75 BNC Adapter Kit 12 4 Accessories and Options ...
Страница 376: ......
Страница 410: ...Markers Figure A 19 Bandwidth Search Example A 34 Basic Measurement Theory ...
Страница 448: ......
Страница 449: ...B Softkey Reference Softkey Reference B 1 ...
Страница 462: ...4Bw Avg5 B 14 Softkey Reference ...
Страница 528: ......
Страница 532: ......
Страница 560: ......