Calibration
Required
for
the
Network
Analyzer
Mode
Step
3:
Measuring
the
Transmission
1.
Connect
a
THRU
between
the
test
port
and
the
return
cable
to
the
analyzer
(connect
to
the
points
at
which
the
test
device
is
connected).
2.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRANSMISSION
.
3.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FWD.
TRANS.
THRU
.
Then
wait
the
S21
frequency
response
is
measured
and
the
softkey
label
is
underlined.
4.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FWD.
MATCH
THRU
.
Then
wait
the
S11
load
match
is
measured
and
the
softkey
label
is
underlined.
5.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRANS.
DONE
.
The
transmission
coecients
are
computed
and
stored.
The
two-port
calibration
menu
is
displayed
(with
the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
TRANSMISSION
softkey
underlined).
Step
4:
Measuring
the
Isolation
1.
If
correction
for
isolation
is
not
required,
press
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNN
ISOLATION
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
OMIT
ISOLATION
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ISOLATION
DONE
.
Skip
to
step
5.
2.
If
correction
for
isolation
is
required,
connect
impedance-matched
LO
ADs
to
the
test
port
and
the
return
port.
3.
Press
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
FWD
ISOL'N
ISOL'N
STD
.
Then
wait
the
S21
isolation
is
measured
and
the
softkey
label
is
underlined.
4.
Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ISOLATION
DONE
.
The
isolation
error
coecients
are
stored.
The
two-port
calibration
menu
is
displayed
(with
the
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNN
ISOLATION
softkey
underlined).
Step
5:
Completing
the
Calibration
1.
Press
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE
2-PORT
CAL
to
complete
the
calibration.
The
error
coecients
are
computed
and
stored.
The
correction
menu
is
displayed
with
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
CORRECTION
ON
off
.
A
corrected
trace
is
displayed.
The
notation
\C2 "
at
the
left
of
the
screen
indicates
that
2-port
error
correction
is
ON.
Step
6:
P
erforming
the
Measurement
1.
Connect
the
test
device
in
the
reverse
direction.
Then
press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRESS
to
CONTINUE
.
2.
Reconnect
the
test
device
in
the
forward
direction.
Then
press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRESS
to
CONTINUE
.
Now
the
error
corrected
trace
is
displayed.
If
you
measure
the
other
test
device
,
press
4
T
rigger
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEASUREMENT
RESTART
.
Then
perform
the
procedure
of
step
6.
Save
the
calibration
data
on
a
oppy
disk
or
memory
disk.
F
or
additional
information
about
calibration,
see
Appendix
A.
Calibration
7-9
Содержание 4395A
Страница 8: ......
Страница 9: ......
Страница 10: ......
Страница 12: ......
Страница 13: ......
Страница 14: ......
Страница 16: ......
Страница 49: ...D 1 Manual Changes by Serial Number D 1 D 2 Manual Changes by Firmware Version D 1 Contents 23 ...
Страница 50: ......
Страница 58: ...Power Requirements Figure 2 1 Power Cable Supplied 2 6 Installation Guide ...
Страница 69: ...Connecting an Active Probe Figure 2 10 Connecting Test Fixture Installation Guide 2 17 ...
Страница 72: ......
Страница 76: ...Step 1 Preparing for the Measurement Figure 3 3 S Parameter Test Set Setup 3 4 Quick Start Guide ...
Страница 93: ...Step 2 Setting Up the 4395A Quick Start Guide 3 21 ...
Страница 130: ......
Страница 144: ......
Страница 154: ......
Страница 198: ...Calibration Required for the Impedance Analyzer Mode Figure 7 2 Connecting Test Fixture 7 16 Calibration ...
Страница 251: ...Measuring Zero Span Figure 8 28 Marker Time Analyzing the Measurement Results 8 49 ...
Страница 262: ......
Страница 278: ......
Страница 298: ...AM Signal Measurement SA where 1Mkr is the 1marker value shown in Figure 10 18 10 20 Examples of Applications ...
Страница 301: ...FM Signal Measurement SA Figure 10 20 Zooming Carrier Signal of FM Signal Examples of Applications 10 23 ...
Страница 374: ...Measurement accessories available 11855A 75 Type N Adapter Kit 11856A 75 BNC Adapter Kit 12 4 Accessories and Options ...
Страница 376: ......
Страница 410: ...Markers Figure A 19 Bandwidth Search Example A 34 Basic Measurement Theory ...
Страница 448: ......
Страница 449: ...B Softkey Reference Softkey Reference B 1 ...
Страница 462: ...4Bw Avg5 B 14 Softkey Reference ...
Страница 528: ......
Страница 532: ......
Страница 560: ......