Chapter 10
157
Command Reference
:CORR:COLL
1
0
. C
o
mm
and
Refe
re
nc
e
:CORR:COLL
Syntax
[:SENSe]:CORRection:COLLect[:ACQuire] {STANdard1|STANdard2|STANdard3}
Description
Measures the compensation data for OPEN/SHORT/LOAD compensation and turns on the
compensation function.
These compensation data are measured for the measurement frequency used in executing
the command (set with the
command).
When using the multi-compensation function (set to ON with the
command), this is measured as the compensation data for the selected channel when
executing the command (selected with the
Parameters
Related commands
Equivalent key
sequence
[Open]
(
[Blue]
-
[4]
) -
OpenMeas
[Short]
(
[Blue]
-
[5]
) -
ShortMeas
[Load]
(
[Blue]
-
[6]
) -
LoadMeas
Description
STANdard1
Specifies the OPEN compensation.
STANdard2
Specifies the SHORT compensation.
STANdard3
Specifies the LOAD compensation.