Chapter 10
145
Command Reference
:CALC:COMP:SEC:LIM
1
0
. C
o
mm
and
Refe
re
nc
e
:CALC:COMP:SEC:LIM
Syntax
:CALCulate:COMParator:SECondary:LIMit <numeric 1>,<numeric 2>
:CALCulate:COMParator:SECondary:LIMit?
Description
Sets the limit range for the secondary parameter used in the comparator function.
This command only sets the limit range. To activate the set limit range, use the
command to enable sorting judgment for the measurement
result of the secondary parameter.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response
{numeric 1},{numeric 2}<newline><^END>
Related commands
Equivalent key
sequence
[Sec Low]
(
[Blue]
-
[Pri Low]
) and
[Sec High]
(
[Blue]
-
[Pri High]
)
[Comprtr]
(
[Blue]
-
[1]
) -
Limit
-
Sec
<numeric 1>
<numeric 2>
Description
The lower limit value of the
limit range.
The upper limit value of the
limit range.
Range
-99.9999E9 to 99.9999E9
-99.9999E9 to 99.9999E9
Initial value
0
0
Unit
Depends on the type of
parameter.
Depends on the type of
parameter.
Содержание 4288A
Страница 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 102: ...100 Chapter8 Avoiding Mistakes Related to Work and Daily Checks Daily Checks executing the self test ...
Страница 212: ...210 AppendixA Manual Changes Manual Changes ...
Страница 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Страница 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Страница 282: ...280 AppendixG 4268A vs 4288A GPIB Command Correspondence Table 4268A vs 4288A GPIB Command Correspondence Table ...
Страница 288: ...286 Index Index ...