:ABOR
4288A GPIB Commands
This section describes the GPIB commands of the 4288A.
:ABOR
Syntax
:ABORt
Description
Resets the trigger system and places the trigger sequence in the idle state.
If the trigger system is set to start up successively (ON is specified with the
command), the trigger system starts up immediately after the transition to the idle state.
For details on the trigger system, refer to “Trigger system” on page 52. (No query)
Related commands
Equivalent key
sequence
No equivalent key is available on the front panel.
:APER
Syntax
[:SENSe][:FIMPedance]:APERture[:MODE] {SHORt|LONG}
[:SENSe][:FIMPedance]:APERture[:MODE]?
Description
Selects the measurement time (integral time) mode.
For information on the specific measurement time of each mode, see “Specification and
Supplemental Performance Characteristics” in the
Operation Manual
.
Parameters
Query response
{SHOR|LONG}<newline><^END>
Equivalent key
sequence
[Meas Time]
Description
SHORt (initial value)
Specifies the short mode.
LONG
Specifies the long mode.
Содержание 4288A
Страница 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 102: ...100 Chapter8 Avoiding Mistakes Related to Work and Daily Checks Daily Checks executing the self test ...
Страница 212: ...210 AppendixA Manual Changes Manual Changes ...
Страница 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Страница 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Страница 282: ...280 AppendixG 4268A vs 4288A GPIB Command Correspondence Table 4268A vs 4288A GPIB Command Correspondence Table ...
Страница 288: ...286 Index Index ...