266
Appendix F
Error Messages
Error number: 21
21
Save failed
Saving the instrument setup into EEPROM has failed.
Although the 4288A will not stop due to this error, the hardware is at fault and needs repair.
Contact an Agilent Technologies sales office or the VAR from which you purchased the
instrument.
22
Recall failed
Recalling the instrument setup from EEPROM has failed.
This error occurs when no instrument setup has been saved in the specified register on
EEPROM. The instrument setup does not change and stays in the state as before executing
the recall.
Confirm that you specified the correct register number when executing the recall.
23
Lockout by handler
Entry using the front panel keys has been disabled through the handler.
You cannot clear this state by using the front panel keys or GPIB command.
Set the /KEY_LOCK signal of the handler interface to HIGH.
25
Correction meas failed
A measurement failure has occurred during measuring the compensation data.
If this error occurs, the compensation data before the measurement remains with no
change.
Confirm that you have made the correct connection for measuring the compensation data
and have performed the compensation procedure correctly.
Содержание 4288A
Страница 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 102: ...100 Chapter8 Avoiding Mistakes Related to Work and Daily Checks Daily Checks executing the self test ...
Страница 212: ...210 AppendixA Manual Changes Manual Changes ...
Страница 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Страница 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Страница 282: ...280 AppendixG 4268A vs 4288A GPIB Command Correspondence Table 4268A vs 4288A GPIB Command Correspondence Table ...
Страница 288: ...286 Index Index ...