Appendix C
Status Reporting System
Status Register Structure
Issuing the
command will clear all bits from the operation status event register.
The 4288A does not support events of the questionable status register. Therefore, all of the
bits in this register are always 0.
Table C-3
Status Bit Definitions of the Operation Status Register
Bit
Position
Name
Description
Condition Register
Event Register
0
Not used
Always 0
Always 0
1
Settling
Set to “1” during the waiting time
to stabilize the measurement
signal.
Set to “1” after the waiting time
needed to stabilize the
measurement signal has elapsed.
2
Ranging
Set to “1” while switching the
measurement range.
Set to “1” when measurement
range switching is completed.
3
Analog Measurement
Set to “1” during analog
measurement
*1
.
Set to “1” when analog
measurement is completed.
4
Measurement
Set to “1” during measurement
*2
.
Set to “1” when measurement is
completed.
5
Waiting for Trigger
Set to “1” while the instrument is
waiting for a trigger
*3
.
Set to “1” when the instrument
starts waiting for a trigger.
6
Not used
Always 0
Always 0
7
Compensation
Set to “1” during compensation
data measurement.
Set to “1” when the compensation
data measurement is completed.
8
Data buffer 1
Set to “1” while data buffer 1 fills.
Set to “1” when data buffer 1 is
full.
9
Data buffer 2
Set to “1” while data buffer 2 fills.
Set to “1” when data buffer 2 is
full.
10
Data buffer 3
Set to “1” while data buffer 3 fills.
Set to “1” when data buffer 3 is
full.
11
Not used
Always 0
Always 0
12
Self-test
Set to “1” during self-test.
Set to “1” when self-test is
completed.
13 to 15
Not used
Always 0
Always 0
*1. This is when the handler interface’s /INDEX signal is active.
*2. This is when the handler interface’s /EOM signal is active.
*3. This is when the trigger system is in trigger wait state. For more information on the trigger
system, refer to “Trigger system” on page 52.
Table C-4
Status Bit Definitions of the Questionable Status Register
Bit
Position
Name
Description
Condition Register
Event Register
0 to 15
Not used
Always 0
Always 0
Содержание 4288A
Страница 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 102: ...100 Chapter8 Avoiding Mistakes Related to Work and Daily Checks Daily Checks executing the self test ...
Страница 212: ...210 AppendixA Manual Changes Manual Changes ...
Страница 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Страница 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Страница 282: ...280 AppendixG 4268A vs 4288A GPIB Command Correspondence Table 4268A vs 4288A GPIB Command Correspondence Table ...
Страница 288: ...286 Index Index ...