32
Chapter 3
Setting Up Measurement Conditions and Display
Sample Program
The details of each part of the program are described below.
Line 70
Sets up the GPIB address.
Lines 90 to 170
Assigns the measurement condition setups (primary parameter: Cp,
secondary parameter: D, measurement signal frequency: 1 kHz,
measurement signal level: 500 mV, measurement range: auto range
mode, measurement time: long mode, cable length: 0 m, averaging
count: 4, and trigger delay time: 1 ms) to the Pri$, Sec$, Freq, Level,
Range$, Meas_time$, Cable, Ave, and Delay variables, respectively.
Lines 190 to 260 Assigns the display setups (result display: ON, the number of display
digits for the measurement result: 5, fixed point display: OFF,
deviation measurement mode for primary parameter: percentage
display, reference value for primary parameter: 10 nF, deviation
measurement mode for secondary parameter: OFF, reference value for
secondary parameter: 0, page number of the instrument setup display
area: 4) to the Disp$, Digit, Fix_msd$, Dev_mode_pri$, Dev_ref_pri,
Dev_mode_sec$, Dev_ref_sec, and Page_no variables, respectively.
Lines 280 to 310
Assigns the measurement signal level monitor setups (current monitor:
OFF, voltage monitor: ON, beep mode setup: FAIL, and save location
number: 9) to I_mon$, V_mon$, Beep_mode$, and Save_no variables,
respectively.
Line 330
Executes a reset.
Lines 340 to 350 Specifies the primary parameter and secondary parameter as Pri$ and
Sec$, respectively.
Lines 360 to 370 Specifies the measurement signal frequency and measurement signal
level as Freq and Level, respectively.
Lines 380 to 430 If Range$ is “Auto,” specifies the measurement range mode as the
auto range. Otherwise, specifies the measurement range mode as the
hold range and the measurement range as Range$.
NOTE
In this example, the measurement range mode is set to the hold mode before the
measurement range is specified. However, in practice, you can omit the setup of the
measurement range mode because setting up the measurement range automatically
specifies the measurement range mode to the hold range.
Lines 440 to 450 Specifies the measurement time and cable length as Meas_time$ and
Cable, respectively.
Lines 460 to 470 Turns ON the averaging function and sets the averaging count to Ave.
Line 480
Specifies the trigger delay time as Delay.
Lines 500 to 510 Specifies the ON/OFF of the result display and the number of display
digits for the measurement result as Disp$ and Digit, respectively.
Lines 520 to 570 If Fix_msd$ is “OFF,” specifies the fixed point display as OFF.
Otherwise, specifies the fixed point display as ON and sets the value
of the highest digit to Fix_msd$.
Содержание 4288A
Страница 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 102: ...100 Chapter8 Avoiding Mistakes Related to Work and Daily Checks Daily Checks executing the self test ...
Страница 212: ...210 AppendixA Manual Changes Manual Changes ...
Страница 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Страница 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Страница 282: ...280 AppendixG 4268A vs 4288A GPIB Command Correspondence Table 4268A vs 4288A GPIB Command Correspondence Table ...
Страница 288: ...286 Index Index ...