![Agilent Technologies 4288A Скачать руководство пользователя страница 90](http://html.mh-extra.com/html/agilent-technologies/4288a/4288a_programming-manual_2867878090.webp)
88
Chapter 7
Sorting Based on Measured Result (Comparator Function)
Reading Out Sort Count Of Each Bin (BIN count function)
Reading Out Sort Count Of Each Bin (BIN count function)
Turning ON/OFF BIN count function
To turn ON/OFF the BIN count function, use the following command.
•
You can count the number of DUTs sorted into each BIN by turning ON the BIN count
function. The maximum value of the count is 999999. If this value is exceeded, the count
does not continue to increase but remains at 999999 (does not return to 0).
When the MULTI compensation function is ON (set to ON with
on page 161), a channel-by-channel count is performed in addition to the normal count
(total of all the channels).
Reading out BIN count value
To read out the BIN count value, use the following commands:
•
•
To read out the BIN count value for each channel when the MULTI compensation function
is ON, use the following commands:
•
•
Clearing (resetting) the BIN count value
To clear all of the BIN count values (initialize to 0), use the following command.
•
Содержание 4288A
Страница 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Страница 102: ...100 Chapter8 Avoiding Mistakes Related to Work and Daily Checks Daily Checks executing the self test ...
Страница 212: ...210 AppendixA Manual Changes Manual Changes ...
Страница 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Страница 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Страница 282: ...280 AppendixG 4268A vs 4288A GPIB Command Correspondence Table 4268A vs 4288A GPIB Command Correspondence Table ...
Страница 288: ...286 Index Index ...