Specifications and Supplemental Performance Characteristics
Supplemental Performance Characteristics
Figure 5-2
Shape of the cap washer
The loading weight applied to the DUT, G [gf], is obtained by substituting the values of t
[mm] into the following equation.
G [gf] = 119 x (3.55 + t ) - 420
NOTE
If a washer with a thickness (t) of less than 1.0 mm is used, measurement may not be
performed accurately.
Table 5-2
Relation between washer thickness and loading weight applied to the DUT
(design values)
t [mm]
Loading weight to the DUT:G [gf]
1.0
120
1.5
180
2.0
240
2.5
300