34
Chapter 3
Operation
Connecting and Measuring DUTs
Connecting and Measuring DUTs
Connect DUTs to the electrodes and take measurements.
CAUTION
Do not connect a DUT, which has an incompatible size.
Step 1.
Remove the cap.
CAUTION
Make sure there is no dirt or other foreign matter in the DUT insertion hole.
Step 2.
Insert the DUT into the insulator hole with tweezers.
Use a magnifying glass to check that the DUT is inserted deeply enough into the insulator
hole for it to contact the bottom electrode.
Figure 3-6
Connecting a DUT
Step 3.
Fit the cap in place with the mark toward the front, and turn it to the right until it is locked.
Step 4.
Take measurements in accordance with the Operation Manual for the measuring instrument
you are using.
Removing the DUT
Use the push ring when removing the DUT.
Содержание 16196A
Страница 7: ...6...
Страница 12: ...Chapter 1 11 Installation Guide Incoming Inspection 1 Installation Guide Figure 1 1 16196A B C D Contents...
Страница 18: ...17 2 Product Overview 2 Product Overview...
Страница 37: ...36 Chapter3 Operation Connecting and Measuring DUTs...
Страница 38: ...37 4 User Maintenance 4 User Maintenance...
Страница 41: ...40 Chapter4 User Maintenance Overview Figure 4 1 Flowchart of User Maintenance...
Страница 59: ...58 Chapter4 User Maintenance Assembling Check...
Страница 68: ...67 6 Service 6 Service This chapter describes the proper maintenance of the fixture and parts replacement...
Страница 85: ...84 Chapter6 Service Assembling Check...