Chapter 3
Operation
Performing Fixture Compensation
Measuring Short Compensation Data
Set the fixture in the short state using the short plate supplied.
Step 1.
Remove the cap. Take out the open plate used to measure the open compensation data.
CAUTION
Make sure there is no dirt or other foreign matter in the DUT insertion hole.
CAUTION
An exclusive type of short plate is supplied with each model. Do not use a short plate from
a different model.
Step 2.
Place the short plate on the insulator assembly with tweezers. Place the rod-shaped
protrusion of the short plate downward, and insert it into the DUT insertion hole.
CAUTION
Handle the short plate with Tweezers. If dirt, etc. gets on it, measuring precision and
repeatability may be adversely affected.
Figure 3-5
Setting the Short State Using the Short Plate
Step 3.
Fit the cap in place with the mark toward the front, and turn it to the right until it is locked.
Step 4.
Take measurements of the data for short compensation in accordance with the Operation
Manual for the measuring instrument you are using.
Содержание 16196A
Страница 7: ...6...
Страница 12: ...Chapter 1 11 Installation Guide Incoming Inspection 1 Installation Guide Figure 1 1 16196A B C D Contents...
Страница 18: ...17 2 Product Overview 2 Product Overview...
Страница 37: ...36 Chapter3 Operation Connecting and Measuring DUTs...
Страница 38: ...37 4 User Maintenance 4 User Maintenance...
Страница 41: ...40 Chapter4 User Maintenance Overview Figure 4 1 Flowchart of User Maintenance...
Страница 59: ...58 Chapter4 User Maintenance Assembling Check...
Страница 68: ...67 6 Service 6 Service This chapter describes the proper maintenance of the fixture and parts replacement...
Страница 85: ...84 Chapter6 Service Assembling Check...