Wear Check
4.
Us
er
Mai
nte
nan
ce
Step 7.
Record the read values as the reference values in the “Check Sheet” (page 50).
Step 8.
Calculate the upper limit value and the lower limit value from the previously set user limits
and the reference values obtained here. Record these in the "Check Sheet".
Step 9.
Tighten the screw on top of the cap loosened in Step 5.
NOTE
In case of the 16196D, please omit the step 9.
Acquisition Procedure (Short Plate Wear Check Reference Value)
Step 1.
Clean the short plate as described in “Cleaning” on page 41.
Step 2.
Remove the cap and place the short plate with the protruding surface down on the insulator
assembly.
Step 3.
Place the cap on the fixture body.
Step 4.
Measure Rs and Ls as described in the Operation Manual for the measuring instrument.
Step 5.
Record the read values as the reference values in the “Check Sheet” (page 51).
Step 6.
Calculate the upper limit value and the lower limit value from the previously set user limits
and the reference values obtained here. Record these in the "Check Sheet".
Содержание 16196A
Страница 7: ...6...
Страница 12: ...Chapter 1 11 Installation Guide Incoming Inspection 1 Installation Guide Figure 1 1 16196A B C D Contents...
Страница 18: ...17 2 Product Overview 2 Product Overview...
Страница 37: ...36 Chapter3 Operation Connecting and Measuring DUTs...
Страница 38: ...37 4 User Maintenance 4 User Maintenance...
Страница 41: ...40 Chapter4 User Maintenance Overview Figure 4 1 Flowchart of User Maintenance...
Страница 59: ...58 Chapter4 User Maintenance Assembling Check...
Страница 68: ...67 6 Service 6 Service This chapter describes the proper maintenance of the fixture and parts replacement...
Страница 85: ...84 Chapter6 Service Assembling Check...