DX-D 600
User Manual
0242F EN 20181218
54
The Ceiling Suspension and the RAD Table are provided also with an Emergency
OFF Switch that cuts the power supply to the Unit. Once pressed, wait almost one
minute before turning ON the Unit again.
IN THE EVENT OF AN EMERGENCY FORCIBLY DEPRESS
THE “EMERGENCY OFF SWITCHES” (USUALLY A RED
MUSHROOM-SHAPED SWITCH) AT ROOM ELECTRICAL
CABINET, CEILING SUSPENSION, AUTO-POSITIONING
CONTROL BOX OR AT THE RAD TABLE. MORE THAN ONE
OF THESE SWITCHES MAY BE PLACED AROUND THE ROOM
FOR GREATER ACCESSIBILITY.
TO ISOLATE THE EQUIPMENT FROM MAINS, TURN OFF THE
SWITCH LOCATED AT THE ROOM ELECTRICAL CABINET.
3.2
SHUTDOWN ROUTINE
TO TURN THE SYSTEM OFF
1.
Turn OFF the NX Workstation.
2.
Press the X-ray Generator Control OFF button, the Green lamp turns
OFF.
3.
Turn OFF the Room Electrical Cabinet Switch.
Do not activate the
Emergency OFF Switch.
4.
The RAD Table, RAD Wall Stand and Ceiling Suspension turn OFF. RAD
Wall Stand ON/OFF Light turns OFF.
AFTER SHUTDOWN OR WHEN THE POWER IS TURNED OFF,
THE BRAKES FOR THE LONGITUDINAL AND TRANSVERSE
AXIS OF THE CEILING SUSPENSION AND THE BRAKES FOR
LONGITUDINAL AND TRANSVERSE MOVEMENT OF THE
TABLETOP ARE RELEASED.
Содержание DX-D 600
Страница 1: ...DX D 600 User Manual Technical Publication 0242F EN 20181218 ...
Страница 2: ......
Страница 20: ...DX D 600 User Manual 0242F EN 20181218 12 This page intentionally left blank ...
Страница 64: ...DX D 600 User Manual 0242F EN 20181218 56 This page intentionally left blank ...
Страница 110: ...DX D 600 User Manual 0242F EN 20181218 102 This page intentionally left blank ...
Страница 126: ...DX D 600 User Manual 0242F EN 20181218 118 This page intentionally left blank ...
Страница 142: ...DX D 600 User Manual 0242F EN 20181218 134 This page intentionally left blank ...
Страница 183: ...DX D 600 User Manual 0242F EN 20181218 175 Illustration 11 9 Ceiling Suspension Specifications Front View ...
Страница 189: ......