DX-D 600
User Manual
0242F EN 20181218
156
Table 8-8
Auto-centering
CODE
USER ERROR MESSAGE
ERROR DESCRIPTION
ACTION
11801
Auto-movement aborted: CS moved
Auto-centering is cancelled because
Ceiling Suspension was moved
Try again
11802
Auto-movement aborted: Wall Stand
moved
Auto-centering is cancelled because
Wall Stand was moved
Try again
11803
Auto-movement aborted: Table moved
Auto-centering is cancelled because
Table was moved
Try again
11804
Auto-movement aborted: Detector
changed
Auto-centering is cancelled because
current Bucky/Detector was changed
Try again
11815
Auto-movement aborted: No Detector
selected
Auto-centering does not start because
NO Bucky/Detector is selected
Select Bucky/Detector and try again
11821
Operative rejected: CS-Longitudinal
Target is out of calibrated range:
Please, check POSITION parameters
Auto-centering does not start because
final position for the Ceiling
Suspension (slave) is out of range on
Longitudinal Axis
Move Bucky/Detector and try again
11822
Auto-movement aborted:
The CS-Longitudinal Axis is blocked:
Please, try again
Auto-centering does not start because
final position for the Ceiling
Suspension (slave) Longitudinal Axis
timed out
Try again
11831
Operative rejected: CS-Transversal
Target is out of calibrated range:
Please, check POSITION parameters
Auto-centering does not start because
final position for the Ceiling
Suspension (slave) is out of range on
Transversal Axis
Move Bucky/Detector and try again
11832
Auto-movement aborted:
The CS-Transversal Axis is blocked:
Please, try again
Auto-centering does not start because
final position for the Ceiling
Suspension (slave) Transversal Axis
timed out
Try again
11841
Operative rejected: CS-Vertical Target
is out of calibrated range: Please,
check POSITION parameters
Auto-centering does not start because
final position for the Ceiling
Suspension (slave) is out of range on
Vertical Axis
Move Bucky/Detector and try again
11842
Auto-movement aborted:
The CS-Vertical Axis is blocked:
Please, try again
Auto-centering does not start because
final position for the Ceiling
Suspension (slave) Vertical Axis timed
out
Try again
Содержание DX-D 600
Страница 1: ...DX D 600 User Manual Technical Publication 0242F EN 20181218 ...
Страница 2: ......
Страница 20: ...DX D 600 User Manual 0242F EN 20181218 12 This page intentionally left blank ...
Страница 64: ...DX D 600 User Manual 0242F EN 20181218 56 This page intentionally left blank ...
Страница 110: ...DX D 600 User Manual 0242F EN 20181218 102 This page intentionally left blank ...
Страница 126: ...DX D 600 User Manual 0242F EN 20181218 118 This page intentionally left blank ...
Страница 142: ...DX D 600 User Manual 0242F EN 20181218 134 This page intentionally left blank ...
Страница 183: ...DX D 600 User Manual 0242F EN 20181218 175 Illustration 11 9 Ceiling Suspension Specifications Front View ...
Страница 189: ......