DX-D 600
User Manual
0242F EN 20181218
162
7.
Select “
Center
” Ion Chamber and Density “
Normal -- 0
”.
Make a RAD exposure and note the exposure mAs and time. For a
proper functioning of the AEC, the exposure must not be aborted by the
AEC back-up timer.
8.
Deselect “Center” and select “Left” Ion Chamber.
Make a RAD exposure and note the exposure mAs and time. For a
proper functioning of the AEC, the exposure must not be aborted by the
AEC back-up timer.
9.
Deselect “Left” and select “Right” Ion Chamber.
Make a RAD exposure and note the exposure mAs and time. For a
proper functioning of the AEC, the exposure must not be aborted by the
AEC back-up timer.
10.
The noted Exposure mAs and time have to be equal
¦
10% between
all three Ion Chambers. If not, contact Service.
11.
Repeat the above steps changing the Density and/or the homogeneous
phantom (e.g. a bucket with 5 cm of water).
Compare the Exposure mAs and time between each Ion Chamber and
between the values noted before (for a lower density or less water, lower
mAs and a shorter time; for half of density or half of water, half of mAs
/ time). If not, contact Service.
12.
Finally, check the proper functioning of the AEC back-up timer by making
a RAD exposure with the selections indicated in step 6., but with the
Collimator blades fully closed.
The exposure must be finished by the AEC back-up timer, that is, the
exposure length is 1.0 second. If not, contact Service.
Содержание DX-D 600
Страница 1: ...DX D 600 User Manual Technical Publication 0242F EN 20181218 ...
Страница 2: ......
Страница 20: ...DX D 600 User Manual 0242F EN 20181218 12 This page intentionally left blank ...
Страница 64: ...DX D 600 User Manual 0242F EN 20181218 56 This page intentionally left blank ...
Страница 110: ...DX D 600 User Manual 0242F EN 20181218 102 This page intentionally left blank ...
Страница 126: ...DX D 600 User Manual 0242F EN 20181218 118 This page intentionally left blank ...
Страница 142: ...DX D 600 User Manual 0242F EN 20181218 134 This page intentionally left blank ...
Страница 183: ...DX D 600 User Manual 0242F EN 20181218 175 Illustration 11 9 Ceiling Suspension Specifications Front View ...
Страница 189: ......