DX-D 600
User Manual
0242F EN 20181218
93
HOW TO DEACTIVATE THE AUTO-TRACKING FUNCTION
Press the AUTO-TRACKING control of the Console or Remote Control,
if available,
press any brake button of the Control Console,
press any button that implies misalignment between the DR Detector and
the Ceiling Suspension, such as selecting a different Workstation, etc.
the auto-tracking function remains active just for 5 minutes after the
Ceiling Suspension becomes inactive. After this time, activate again the
Auto-tracking function.
When working with the Programmed Positions (PP) remember that the
Auto-tracking function will be activated or not, depending on the last PP which
has been performed. If the Programmed position was set to have the
Auto-tracking ON, it remains on once this position is performed. If the position
was configured to have the Auto-tracking off, it must be activated on the
Console to turn it on.
4.2.3.8 AUTO-POSITIONING
Auto-positioning assists the Operator with the Ceiling Suspension movements,
and also with the RAD Table and RAD Wall Stand DR. Then, the operator must
adjust the position as needed for the exposure, if required.
The Ceiling Suspension keeps different positions.
In case of the Automatic Ceiling Suspension refer to
Section 4.2.4
“Position. Automatic System Auto-positioning“
and
Section 4.2.6 -
“Programmed Positions (PP)”.
In case of the Semi-automatic Ceiling Suspension refer to
Section 4.2.5
“Position. Semi-automatic System Auto-positioning“
and
Section 4.2.6
- “Programmed Positions (PP)”.
Содержание DX-D 600
Страница 1: ...DX D 600 User Manual Technical Publication 0242F EN 20181218 ...
Страница 2: ......
Страница 20: ...DX D 600 User Manual 0242F EN 20181218 12 This page intentionally left blank ...
Страница 64: ...DX D 600 User Manual 0242F EN 20181218 56 This page intentionally left blank ...
Страница 110: ...DX D 600 User Manual 0242F EN 20181218 102 This page intentionally left blank ...
Страница 126: ...DX D 600 User Manual 0242F EN 20181218 118 This page intentionally left blank ...
Страница 142: ...DX D 600 User Manual 0242F EN 20181218 134 This page intentionally left blank ...
Страница 183: ...DX D 600 User Manual 0242F EN 20181218 175 Illustration 11 9 Ceiling Suspension Specifications Front View ...
Страница 189: ......