Table 8-2
Ceiling Suspension Motor Controller Boards
CODE
USER ERROR MESSAGE
ERROR DESCRIPTION
ACTION
11100
Control Board Error: CS-Longitudinal
Motor Controller Board not connected
or I2C communication problems
Check the Motor Controller I/F
Connection
11101
Control Board Error: CS-Transversal
Motor Controller Board not connected
or I2C communication problems
Check the Motor Controller I/F
Connection
11102
Control Board Error: CS-Vertical
Motor Controller Board not connected
or I2C communication problems
Check the Motor Controller I/F
Connection
11103
Control Board Error: CS-Angular
Motor Controller Board not connected
or I2C communication problems
Check the Motor Controller I/F
Connection
11104
Control Board Error: CS-Rotational
Motor Controller Board not connected
or I2C communication problems
Check the Motor Controller I/F
Connection
11105
Invalid Calibration Data:
CS-Longitudinal
Axis not calibrated,
calibration file corrupted
Calibrate once again
If message persists change SBC/CF
11106
Invalid Calibration Data:
CS-Transversal
Axis not calibrated,
calibration file corrupted
Calibrate once again
If message persists change SBC/CF
11107
Invalid Calibration Data: CS-Vertical
Axis not calibrated,
calibration file corrupted
Calibrate once again
If message persists change SBC/CF
11108
Invalid Calibration Data: CS-Angular
Axis not calibrated,
calibration file corrupted
Calibrate once again.
If message persists change SBC/CF
11109
Invalid Calibration Data: CS-Rotational
Axis not calibrated,
calibration file corrupted
Calibrate once again
If message persists change SBC/CF
11110
Operative rejected: CS-Longitudinal
position out of calibrated range:
Right there, only MANUAL
movement allowed
Axis moved out of
defined range manually.
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
Check position feedback
Contact Service
11111
Board failure: CS-Longitudinal
No motor power or
motor/driver defective.
Potentiometer/Tachometer or
circuitry defective.
Coupling did not engage
Change board
11112
End-switch limit: CS-Longitudinal
Axis moved out of
defined range manually.
Axis not calibrated.
Potentiometer misaligned.
Potentiometer or Potentiometer
Circuitry defective.
A/D converter problem
Check End Switch Limit position,
activation, operation.
Check End Switch Limit range
Содержание DX-D 600
Страница 1: ...DX D 600 User Manual Technical Publication 0242F EN 20181218 ...
Страница 2: ......
Страница 20: ...DX D 600 User Manual 0242F EN 20181218 12 This page intentionally left blank ...
Страница 64: ...DX D 600 User Manual 0242F EN 20181218 56 This page intentionally left blank ...
Страница 110: ...DX D 600 User Manual 0242F EN 20181218 102 This page intentionally left blank ...
Страница 126: ...DX D 600 User Manual 0242F EN 20181218 118 This page intentionally left blank ...
Страница 142: ...DX D 600 User Manual 0242F EN 20181218 134 This page intentionally left blank ...
Страница 183: ...DX D 600 User Manual 0242F EN 20181218 175 Illustration 11 9 Ceiling Suspension Specifications Front View ...
Страница 189: ......