RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE
AND MOBILE RF COMMUNICATIONS EQUIPMENT AND THE X-RAY SYSTEM
(IEC 60601-1-2:2007)
This DX-D 600 Radiographic Room is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of this DX-D 600 Radiographic Room can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the DX-D 600
Radiographic Room as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
Separation distance according to frequency of transmitter
m
Rated maximum output power of transmitter
W
150 KHz to 80 MHz
d
=
1.2
P
80 MHz to 800 MHz
d
=
1.2
P
800 MHz to 2.5 GHz
d
=
2.3
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
TYPICAL RF DEVICES (Worst-Case scenario)
Device: Power @ Frequency
Recommended
distance(m)
GMRS device (Professional Walkie--Talkie): 5 W @ 462--467 MHz
2.7
GSM / UMTS cell phone: 2 W @ 850/1700/1900 MHz
3.3
FRS device (Amateur Walkie--Talkie): 500 mW @ 462--467 MHz
0.9
WiFi / Bluetooth devices: 100 mW @ 2400--2500 MHz
0.8
DECT devices (modern cordless phones): 100mW @ 1880--1900 MHz
0.8
RFID reader (3): 10 mW @ 125--150 KHz / 13.56 MHz
0.12
RFID reader (3): 10 mW @ 902--928 MHz / 2400--2500 MHz
0.23
Station transmitter ATSC TV broadcasting: 100 kW @ 54--800 MHz
380
Station transmitter ATSC TV broadcasting: 100 kW @ 800--890 MHz
730
Station transmitter FM radio broadcasting: 100 kW @ 87.5--108 MHz
380
For transmitters rated at a maximum output power not listed above, the recommended separation distance ’d’ in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where ’P’ is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1 -- At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 -- These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Содержание DX-D 600
Страница 1: ...DX D 600 User Manual Technical Publication 0242F EN 20181218 ...
Страница 2: ......
Страница 20: ...DX D 600 User Manual 0242F EN 20181218 12 This page intentionally left blank ...
Страница 64: ...DX D 600 User Manual 0242F EN 20181218 56 This page intentionally left blank ...
Страница 110: ...DX D 600 User Manual 0242F EN 20181218 102 This page intentionally left blank ...
Страница 126: ...DX D 600 User Manual 0242F EN 20181218 118 This page intentionally left blank ...
Страница 142: ...DX D 600 User Manual 0242F EN 20181218 134 This page intentionally left blank ...
Страница 183: ...DX D 600 User Manual 0242F EN 20181218 175 Illustration 11 9 Ceiling Suspension Specifications Front View ...
Страница 189: ......