124
Illustration 6-5
DR Detector Tilting Assembly
Tilting Brake
90o Position Lock
0o Position Lock
1.
Unlock the 0
o
/90
o
Detent.
2.
Press and hold one of the Tilting Brake pushbuttons located on the DR
Detector Support.
3.
Tilt manually the DR Detector up to the desired position.
4.
The angle is indicated on the Tilting Goniometer on the right lateral.
5.
Release the pushbutton, to get the DR Detector position locked.
6.
If final position is 0
o
or 90
o
insert the Lock Detent.