background image

17

USA

9.

Troubleshooting list

Have defective products repaired by Aesculap Technical Service, see 
Chapter 10.

9.1

Display error messages

Faults that are recognized by the control unit are shown as error messages 
in the display.

There are three types of error messages:

System error (text in red field): It is not possible to work with the con-
trol unit or system.

Accessory error (text in yellow field): It is possible to work with a dif-
ferent component.

Operating error (text in blue field): It is possible to work with the sys-
tem after rectifying the cause of the error.

Note

Many faults cannot be clearly categorized. They can be caused by an oper-
ating error or an accessory error. In such cases, an operating error is 
assumed first, to avoid any unnecessary changing or sending in of new 
products.

9.1.1

System error

9.1.2

Accessory error

Message displayed

Cause

Remedy

System error
Switch the control unit off and on.
If the error re-occurs, replace the control unit.

The internal monitoring of the control unit 
recognizes an error or fault.

Switch the control unit off and on again.
If the message appears again: Replace the control 
unit.

Message displayed

Cause

Remedy

Motor or motor cable could be damaged

Motor cable or applied part defective

Replace motor cable or applied part.

Malfunction of foot control
If this message is displayed frequently, please per-
form a technical inspection on the foot control

Foot control defective

Replace foot control.

Critical Foot Control Error
Unexpected error occurred in the area of the foot 
control. If this message does not disappear auto-
matically, please restart the device. 

Foot control defective

Replace foot control.

Critical error in manual control module
Unexpected error occurred in the area of the hand 
control. If this occurs repeatedly, please replace 
the hand control.

Unexpected error in the area of the hand con-
trol 

Restart device

Hand control defective

Replace hand control.

Prohibited combination of accessories
A prohibited combination of motor and cable cod-
ing was found on socket 1/socket 2.

A prohibited combination of application part 
and motor cable is connected to the control 
unit.

Check inserted products and replace with a per-
missible combination of application part and 
motor cable as applicable.

Motor not recognized
It was impossible to recognize the motor type. 
Please replace the motor. If this does not work, 
replace the motor cable as well.

Wrong applied part
Applied part defective

Change applied part.

Motor cable defective

Replace motor cable.

Unknown motor type
Please verify that the software version of the 
device is suitable for this motor type.

Software version of the control unit is incom-
patible with the application part used.

Verify software version of the control unit.
Have control unit software updated by Technical 
Service.

Motor overheated
The motor currently being used is overheating.
Please let the motor cool down or use another suit-
able motor.

Motor of the applied part has overheated

Allow the applied part to cool down.
If the applied part overheats excessively: Replace 
the applied part.

Motor blocked
Stop motor actuation and release blockage
If the error re-occurs, replace the product.

Applied part blocked

Stop applied part actuation and release blockage.
If the error occurs when the applied part is running 
idle: Replace the applied part.

Document No.:  IFU   -  Version: 3.0  -  Document ID: SOP-AIC-5001568   Date/Time Printed/Viewed:  2022-04-13  15:59 (CET)

Effective

Содержание ELAN 4

Страница 1: ...ESCULAP USA Instructions for use Technical description ELAN 4 electro control unit GA800 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 13 15 59 CET Effective...

Страница 2: ...11 12 13 14 11 15 16 18 17 1 2 3 4 5 6 7 8 9 10 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 13 15 59 CET Effective...

Страница 3: ...19 20 21 22 23 24 25 27 28 29 30 26 29 28 30 31 GA810 GA808 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 13 15 59 CET Effective...

Страница 4: ...12804 GA395SU GA808 GA849 GA822 GA861 GA864 GA865 GA869 GA860 GA836 GA831 GA837 GA832 GA824 GA833 GA844 GA810 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 1...

Страница 5: ...tton 29 Pedal 30 Rotational direction button for motor 31 Speed range limit button only for wireless foot control GA810 Symbols on product and packaging Caution Observe important safety information su...

Страница 6: ...g L10 GA867 ELAN 4 electro Standard handpiece 2 ring L13 GA868 ELAN 4 electro Standard handpiece 2 ring L17 GA869 ELAN 4 electro Standard handpiece 2 ring L22 Drill GA844 ELAN 4 electro power drill Sy...

Страница 7: ...eed up Slow down Oscillation mode Oscillation angle Flow rate Increase the value Reduce the value System volume Display brightness System language Reset device to factory settings Go back to menu stru...

Страница 8: ...control unit 11 6 2 Function checks 11 6 3 Safe operation 11 6 3 1 Activating the motor pump control panel 12 6 3 2 Changing the maximum speed setting rotational direction of the applied part 12 6 3 3...

Страница 9: ...al circulatory system Note The safe and effective use of electrically driven applied parts depends heav ily on influences that only the user can control him or herself Therefore the specifications pro...

Страница 10: ...ferent applied parts and has one connection socket for a foot control Only one applied part can be operated at a time Note The essential performance features are defined by the speed and rota tional d...

Страница 11: ...t the device via the potential equalization terminal at the rear panel of the device to the potential equalization system of the room used for medical purposes Note The potential equalization lead can...

Страница 12: ...n of the product from the main power supply is only guaranteed when the power cord is unplugged Switching off the product Activate the power OFF switch 8 Unplug the power cord from the power socket 13...

Страница 13: ...mp tube does not get clamped in Insert the bottle holder 16 for the sterile fluid bottle in bottle holder mount 15 Insert the punching spike into the sterile fluid bottle When using sterile fluid bott...

Страница 14: ...o be used are in good working order Check the secure connection of all products to be used Ensure that the parameters are set and the unit is operated according to the instructions for use and the saf...

Страница 15: ...3 3 Activating the coolant and irrigation pump Flush function continuous flushing Activate the pump control panel see Chapter 6 3 1 Activate the coolant and irrigation pump Activate the button for sw...

Страница 16: ...wise rotation on the control unit The motor direction indicator for counterclockwise rotation is illumi nated in the applied part control panel Activate the control element of the hand control The app...

Страница 17: ...les c Press button import or export Follow the instructions on the display Note Profiles must be saved after creation or modification Save Profile Press button 6 4 2 Motor settings The motor settings...

Страница 18: ...l number of the wireless foot control and select Yes to con firm the message Disconnect the GA810 wireless foot control by pressing the Disconnecting from GA810 button or switch off the control unit f...

Страница 19: ...nfecting agents Only use cleaning disinfecting agents approved for surface clean ing Follow the manufacturer s instructions for the respective cleaning disinfecting agent Do not clean the product in a...

Страница 20: ...rol defective Replace foot control Critical error in manual control module Unexpected error occurred in the area of the hand control If this occurs repeatedly please replace the hand control Unexpecte...

Страница 21: ...test Release the activation If the pedal or button is not activated the foot control is defective If necessary replace foot control Multiple control elements of the foot control acti vated Multiple co...

Страница 22: ...applied part Coupling processes cannot be per formed Release the release button on the motor cable between connecting dis connecting and releasing locking the applied part and activate again Motor cou...

Страница 23: ...art operated in counter clockwise rotation Serrated tool run in counterclock wise rotation mode Operate serrated tool in clockwise rotation No central alignment of the crani otomy reamer to the dura p...

Страница 24: ...ap Repair Hotline Phone 1 800 214 3392 Fax 1 314 895 4420 Other service addresses can be obtained from the address indicated above 11 Accessories Spare parts 11 1 ELAN 4 electro motor cables applied p...

Страница 25: ...piece IIa GA861 ELAN 4 electro Standard handpiece 1 ring L4 IIa GA862 ELAN 4 electro Standard handpiece 1 ring L7 IIa GA863 ELAN 4 electro Standard handpiece 1 ring L10 IIa GA864 ELAN 4 electro Stand...

Страница 26: ...Operation Storage and transport Temperature 10 C to 40 C 10 C to 50 C Relative humidity 30 to 75 10 to 90 Atmospheric pressure 700 hPa to 1 060 hPa 500 hPa to 1 060 hPa The recycling pass can be downl...

Страница 27: ...n Germany Phone 49 0 7461 95 0 Fax 49 0 7461 95 26 00 www bbraun com AESCULAP a B Braun brand SOP AIC 5001568 TA014401 2021 07 Rev 3 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time P...

Отзывы: