4.3.4

Applied part types

The control unit recognizes the different types of applied parts (motors and 
handpieces). These are shown in the display as a combination of symbol 
and text. The size and display depend on which socket the applied part is 
connected to and whether the applied part is active or locked.

4.3.5

Identification of control elements on system components

Control elements on the system components of the ELAN 4 electro motor 
system are labeled with a gold marking.

4.3.6

Recognition of connected motor cable and applied part 

The settings (upper speed limit, pump status and delivery volume) last 
selected with this application part type on this bushing are retrieved upon 
inserting the same application part type.

When several different hand/foot controls are connected to the control 
unit, they are prioritized in the following order (higher items have priority 
over lower):

Hand controls GA804/GA805 or drill tool GA844

Foot control GA808

GA810 wireless foot control

4.3.7

Overload protection

To protect the micro motors in the applied parts against damage from 
overheating, the motor temperature is monitored. If the temperature is too 
high a warning signal sounds and a thermometer symbol appears in the 
display 

2

.

If the temperature continues to be too high, the applied part is switched 
off. The following message appears in the display 

2

: “The motor currently 

being used is overheating. Please let the motor cool down or use another 
suitable motor”.

After a cooling phase the applied part can be reactivated.

It is recommended to have a second applied part ready at hand.

4.3.8

Coolant and irrigation pump

The control unit is equipped with a coolant and irrigation pump 

3

.

The pump can be activated either in the corresponding pump control panel 
or via the corresponding button on the foot control. It can be started by 
activating the motor or via the “Flush” function (continuous flushing). The 
delivery rate can only be adjusted via the pump control panel.

5.

Preparation and setup

Non-compliance with the following instructions will preclude all respon-
sibility and liability in this respect on the part of Aesculap.

When setting up and operating the product, adhere to

– national regulations for installation and operation,

– national regulations on fire and explosion protection.

Note

For the safety of patients and users it is essential that the mains power cord 
and, especially, the protective earth connection are intact. In many cases 
defective or missing protective earth connections are not registered imme-
diately.

Connect the device via the potential equalization terminal at the rear 
panel of the device to the potential equalization system of the room 
used for medical purposes.

Note

The potential equalization lead can be ordered from the manufacturer as 
art. no. GK535 (4 m length) or TA008205 (0.8 m length).

5.1

Set-up environment/set-up location

 

DANGER

Fire and explosion hazard!

Use the product outside of explosion hazard zones (e.g. areas with 
highly pure oxygen or anesthetic gases).

The ELAN 4 electro control unit is authorized for operation in the operat-
ing room.

Note

After being installed and put into operation, the control unit must not be 
transported or moved to a different set-up location.

Note

The control unit must not be placed on an Aesculap mobile stand (GA415, 
GA416 and GD416M).

Make certain that the ventilation slots in the bottom of the housing 
and in the rear panel of the control unit are not covered, e.g. by an OR 
cloth.

Make certain that the control elements, mains power switch and power 
socket 

13

 are freely accessible for the user.

Make certain the system is set up on a sufficiently stable support (e.g. 
a table, ceiling support, equipment cart, etc.).

Follow the instructions for use of the support.

5.2

Stacking of units

Do not exceed a stack height of 415 mm.

Place the units in a stable position.

Aesculap devices must be stacked on top of each other facing identi-
cally.

Never move stack devices.

6.

Working with the ELAN 4 electro sys-
tem

6.1

System set-up

 

WARNING

Risk of injury and material damage due to inadvertent activation of 
the applied part!

Secure applied parts that are not actually used against inadvertent 
activation ("Off" position), see section on locking applied part 
("Off" position).

6.1.1

Connecting the accessories

Combinations of accessories that are not mentioned in the present 
instructions for use may only be employed if they are specifically intended 
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.

Also note that any equipment connected at the interfaces must demon-
strably meet the respective IEC standards (e.g. IEC 60950 for data process-
ing equipment, IEC/DIN EN 60601-1 for electromedical devices). 

All configurations must fulfill the basic standard IEC/DIN EN 60601-1. Any 
individual connecting devices with one another is responsible for such 
configuration and must ensure compliance with basic standard IEC/DIN 
EN 60601-1 or applicable national standards.

Follow the instructions for use of individual accessories.

In case of questions, contact your B. Braun/Aesculap partner or 
Aesculap Technical Service, Address see Chapter 10.

Document No.:  IFU   -  Version: 3.0  -  Document ID: SOP-AIC-5001568   Date/Time Printed/Viewed:  2022-04-13  15:59 (CET)

Effective

Содержание ELAN 4

Страница 1: ...ESCULAP USA Instructions for use Technical description ELAN 4 electro control unit GA800 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 13 15 59 CET Effective...

Страница 2: ...11 12 13 14 11 15 16 18 17 1 2 3 4 5 6 7 8 9 10 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 13 15 59 CET Effective...

Страница 3: ...19 20 21 22 23 24 25 27 28 29 30 26 29 28 30 31 GA810 GA808 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 13 15 59 CET Effective...

Страница 4: ...12804 GA395SU GA808 GA849 GA822 GA861 GA864 GA865 GA869 GA860 GA836 GA831 GA837 GA832 GA824 GA833 GA844 GA810 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 1...

Страница 5: ...tton 29 Pedal 30 Rotational direction button for motor 31 Speed range limit button only for wireless foot control GA810 Symbols on product and packaging Caution Observe important safety information su...

Страница 6: ...g L10 GA867 ELAN 4 electro Standard handpiece 2 ring L13 GA868 ELAN 4 electro Standard handpiece 2 ring L17 GA869 ELAN 4 electro Standard handpiece 2 ring L22 Drill GA844 ELAN 4 electro power drill Sy...

Страница 7: ...eed up Slow down Oscillation mode Oscillation angle Flow rate Increase the value Reduce the value System volume Display brightness System language Reset device to factory settings Go back to menu stru...

Страница 8: ...control unit 11 6 2 Function checks 11 6 3 Safe operation 11 6 3 1 Activating the motor pump control panel 12 6 3 2 Changing the maximum speed setting rotational direction of the applied part 12 6 3 3...

Страница 9: ...al circulatory system Note The safe and effective use of electrically driven applied parts depends heav ily on influences that only the user can control him or herself Therefore the specifications pro...

Страница 10: ...ferent applied parts and has one connection socket for a foot control Only one applied part can be operated at a time Note The essential performance features are defined by the speed and rota tional d...

Страница 11: ...t the device via the potential equalization terminal at the rear panel of the device to the potential equalization system of the room used for medical purposes Note The potential equalization lead can...

Страница 12: ...n of the product from the main power supply is only guaranteed when the power cord is unplugged Switching off the product Activate the power OFF switch 8 Unplug the power cord from the power socket 13...

Страница 13: ...mp tube does not get clamped in Insert the bottle holder 16 for the sterile fluid bottle in bottle holder mount 15 Insert the punching spike into the sterile fluid bottle When using sterile fluid bott...

Страница 14: ...o be used are in good working order Check the secure connection of all products to be used Ensure that the parameters are set and the unit is operated according to the instructions for use and the saf...

Страница 15: ...3 3 Activating the coolant and irrigation pump Flush function continuous flushing Activate the pump control panel see Chapter 6 3 1 Activate the coolant and irrigation pump Activate the button for sw...

Страница 16: ...wise rotation on the control unit The motor direction indicator for counterclockwise rotation is illumi nated in the applied part control panel Activate the control element of the hand control The app...

Страница 17: ...les c Press button import or export Follow the instructions on the display Note Profiles must be saved after creation or modification Save Profile Press button 6 4 2 Motor settings The motor settings...

Страница 18: ...l number of the wireless foot control and select Yes to con firm the message Disconnect the GA810 wireless foot control by pressing the Disconnecting from GA810 button or switch off the control unit f...

Страница 19: ...nfecting agents Only use cleaning disinfecting agents approved for surface clean ing Follow the manufacturer s instructions for the respective cleaning disinfecting agent Do not clean the product in a...

Страница 20: ...rol defective Replace foot control Critical error in manual control module Unexpected error occurred in the area of the hand control If this occurs repeatedly please replace the hand control Unexpecte...

Страница 21: ...test Release the activation If the pedal or button is not activated the foot control is defective If necessary replace foot control Multiple control elements of the foot control acti vated Multiple co...

Страница 22: ...applied part Coupling processes cannot be per formed Release the release button on the motor cable between connecting dis connecting and releasing locking the applied part and activate again Motor cou...

Страница 23: ...art operated in counter clockwise rotation Serrated tool run in counterclock wise rotation mode Operate serrated tool in clockwise rotation No central alignment of the crani otomy reamer to the dura p...

Страница 24: ...ap Repair Hotline Phone 1 800 214 3392 Fax 1 314 895 4420 Other service addresses can be obtained from the address indicated above 11 Accessories Spare parts 11 1 ELAN 4 electro motor cables applied p...

Страница 25: ...piece IIa GA861 ELAN 4 electro Standard handpiece 1 ring L4 IIa GA862 ELAN 4 electro Standard handpiece 1 ring L7 IIa GA863 ELAN 4 electro Standard handpiece 1 ring L10 IIa GA864 ELAN 4 electro Stand...

Страница 26: ...Operation Storage and transport Temperature 10 C to 40 C 10 C to 50 C Relative humidity 30 to 75 10 to 90 Atmospheric pressure 700 hPa to 1 060 hPa 500 hPa to 1 060 hPa The recycling pass can be downl...

Страница 27: ...n Germany Phone 49 0 7461 95 0 Fax 49 0 7461 95 26 00 www bbraun com AESCULAP a B Braun brand SOP AIC 5001568 TA014401 2021 07 Rev 3 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time P...

Отзывы: