background image

19

20

21

22

23

24

25

27

28

29

30

26

29

28

30

31

GA810

GA808

Document No.:  IFU   -  Version: 3.0  -  Document ID: SOP-AIC-5001568   Date/Time Printed/Viewed:  2022-04-13  15:59 (CET)

Effective

Содержание ELAN 4

Страница 1: ...ESCULAP USA Instructions for use Technical description ELAN 4 electro control unit GA800 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 13 15 59 CET Effective...

Страница 2: ...11 12 13 14 11 15 16 18 17 1 2 3 4 5 6 7 8 9 10 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 13 15 59 CET Effective...

Страница 3: ...19 20 21 22 23 24 25 27 28 29 30 26 29 28 30 31 GA810 GA808 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 13 15 59 CET Effective...

Страница 4: ...12804 GA395SU GA808 GA849 GA822 GA861 GA864 GA865 GA869 GA860 GA836 GA831 GA837 GA832 GA824 GA833 GA844 GA810 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time Printed Viewed 2022 04 1...

Страница 5: ...tton 29 Pedal 30 Rotational direction button for motor 31 Speed range limit button only for wireless foot control GA810 Symbols on product and packaging Caution Observe important safety information su...

Страница 6: ...g L10 GA867 ELAN 4 electro Standard handpiece 2 ring L13 GA868 ELAN 4 electro Standard handpiece 2 ring L17 GA869 ELAN 4 electro Standard handpiece 2 ring L22 Drill GA844 ELAN 4 electro power drill Sy...

Страница 7: ...eed up Slow down Oscillation mode Oscillation angle Flow rate Increase the value Reduce the value System volume Display brightness System language Reset device to factory settings Go back to menu stru...

Страница 8: ...control unit 11 6 2 Function checks 11 6 3 Safe operation 11 6 3 1 Activating the motor pump control panel 12 6 3 2 Changing the maximum speed setting rotational direction of the applied part 12 6 3 3...

Страница 9: ...al circulatory system Note The safe and effective use of electrically driven applied parts depends heav ily on influences that only the user can control him or herself Therefore the specifications pro...

Страница 10: ...ferent applied parts and has one connection socket for a foot control Only one applied part can be operated at a time Note The essential performance features are defined by the speed and rota tional d...

Страница 11: ...t the device via the potential equalization terminal at the rear panel of the device to the potential equalization system of the room used for medical purposes Note The potential equalization lead can...

Страница 12: ...n of the product from the main power supply is only guaranteed when the power cord is unplugged Switching off the product Activate the power OFF switch 8 Unplug the power cord from the power socket 13...

Страница 13: ...mp tube does not get clamped in Insert the bottle holder 16 for the sterile fluid bottle in bottle holder mount 15 Insert the punching spike into the sterile fluid bottle When using sterile fluid bott...

Страница 14: ...o be used are in good working order Check the secure connection of all products to be used Ensure that the parameters are set and the unit is operated according to the instructions for use and the saf...

Страница 15: ...3 3 Activating the coolant and irrigation pump Flush function continuous flushing Activate the pump control panel see Chapter 6 3 1 Activate the coolant and irrigation pump Activate the button for sw...

Страница 16: ...wise rotation on the control unit The motor direction indicator for counterclockwise rotation is illumi nated in the applied part control panel Activate the control element of the hand control The app...

Страница 17: ...les c Press button import or export Follow the instructions on the display Note Profiles must be saved after creation or modification Save Profile Press button 6 4 2 Motor settings The motor settings...

Страница 18: ...l number of the wireless foot control and select Yes to con firm the message Disconnect the GA810 wireless foot control by pressing the Disconnecting from GA810 button or switch off the control unit f...

Страница 19: ...nfecting agents Only use cleaning disinfecting agents approved for surface clean ing Follow the manufacturer s instructions for the respective cleaning disinfecting agent Do not clean the product in a...

Страница 20: ...rol defective Replace foot control Critical error in manual control module Unexpected error occurred in the area of the hand control If this occurs repeatedly please replace the hand control Unexpecte...

Страница 21: ...test Release the activation If the pedal or button is not activated the foot control is defective If necessary replace foot control Multiple control elements of the foot control acti vated Multiple co...

Страница 22: ...applied part Coupling processes cannot be per formed Release the release button on the motor cable between connecting dis connecting and releasing locking the applied part and activate again Motor cou...

Страница 23: ...art operated in counter clockwise rotation Serrated tool run in counterclock wise rotation mode Operate serrated tool in clockwise rotation No central alignment of the crani otomy reamer to the dura p...

Страница 24: ...ap Repair Hotline Phone 1 800 214 3392 Fax 1 314 895 4420 Other service addresses can be obtained from the address indicated above 11 Accessories Spare parts 11 1 ELAN 4 electro motor cables applied p...

Страница 25: ...piece IIa GA861 ELAN 4 electro Standard handpiece 1 ring L4 IIa GA862 ELAN 4 electro Standard handpiece 1 ring L7 IIa GA863 ELAN 4 electro Standard handpiece 1 ring L10 IIa GA864 ELAN 4 electro Stand...

Страница 26: ...Operation Storage and transport Temperature 10 C to 40 C 10 C to 50 C Relative humidity 30 to 75 10 to 90 Atmospheric pressure 700 hPa to 1 060 hPa 500 hPa to 1 060 hPa The recycling pass can be downl...

Страница 27: ...n Germany Phone 49 0 7461 95 0 Fax 49 0 7461 95 26 00 www bbraun com AESCULAP a B Braun brand SOP AIC 5001568 TA014401 2021 07 Rev 3 Document No IFU Version 3 0 Document ID SOP AIC 5001568 Date Time P...

Отзывы: