18
USA
9.1.3
Operating error
Message displayed
Cause
Remedy
Motor turned off in operation.
The motor was turned to the off-position n opera-
tion.
Stop the engine before turning off the engine
The motor was turned to the off-position n
operation.
Stop the engine before turning off the engine
Motor not recognized
Switch the motor to "Off" position and then back
to "On" position
Applied part connected to control unit in "On"
position
Lock the applied part ("Off" position).
The control unit recognizes the applied part type.
In order to work: Release the applied part ("On"
position).
Two motors simultaneously in on position
Please move one of the motors to the Off position.
Foot control is activated while two applied
parts are released on the motor cable ("On"
position)
Note: Only one applied part can be operated at
a time.
Only release the applied part intended to be oper-
ated ("On" position).
Lock the applied part which isn't to be operated
("Off" position).
Activation in off-position
Activation of the motor in Off position
Switch the motor into the On position before acti-
vation.
Foot control is activated while the applied part
is locked on the motor cable ("Off" position)
Release the applied part ("On" position).
Activation of foot control without motor
Connect a motor to the control unit.
Foot control is activated while the applied part
is not connected to the control unit
Connect the motor cable to the control unit.
Connect the applied part to the motor cable.
Activation of foot control during initialization
The pedal or a foot control button was activated
during initialization.
Please release pedal and buttons.
If this message still does not disappear, please sub-
ject the foot control to a technical check.
Pedal or button on the foot control is activated
during automatic self-test.
Release the activation.
If the pedal or button is not activated, the foot
control is defective.
If necessary, replace foot control.
Multiple control elements of the foot control acti-
vated
Multiple control elements were activated simulta-
neously.
Please release all control elements.
If this message does not disappear, please have the
foot control inspected.
Multiple control elements were activated
simultaneously.
Release all foot control operating elements.
If this message continues to appear: foot control
must undergo a technical inspection.
Extended button activation on foot control
A foot control button was activated for longer than
20 seconds without interruption.
Please release all buttons.
If this message does not disappear, please have the
foot control inspected.
A foot control operating element was acti-
vated for longer than 20 seconds without
interruption.
Release all foot control operating elements.
If this message continues to appear: foot control
must undergo a technical inspection.
Activation of manual control without motor
Connect a motor to the control unit.
Manual control was activated without an
application part being connected to the con-
trol unit.
Connect application part to control unit.
Manual control activated upon connecting
Please move the manual control lever/slider to an
unactivated position one time.
Manual control lever/slider was activated dur-
ing connection.
Move manual control level/slider to an unactivated
position one time.
Touchscreen timeout
The touchscreen was activated for longer than
20 seconds without interruption.
Please release the touchscreen.
If this message continues to display, the device
may be damaged.
The touchscreen was activated for longer than
20 seconds without interruption.
Release touchscreen.
If this message continues to display, the control
unit may be damaged. Replace the control unit.
Document No.: IFU - Version: 3.0 - Document ID: SOP-AIC-5001568 Date/Time Printed/Viewed: 2022-04-13 15:59 (CET)