16
USA
7.2
Reusable products
The service life of the product is limited by damage, normal wear, type and
duration of the application, handling, storage and transportation of the
product.
Careful visual and functional testing prior to next use is the best way to
identify a malfunctioning product.
7.3
Preparation before cleaning
►
Separate the products from each other immediately after use.
►
Shut down control unit
1
immediately after use, see Chapter 6.1.5.
►
Remove any visible surgical residues as much as possible with a damp,
lint-free cloth.
7.4
Cleaning/disinfection
7.4.1
Product-specific safety instructions for the reprocessing pro-
cedure
DANGER
Risk of electric shock and fire hazard!
►
Unplug the device before cleaning.
►
Do not use flammable or explosive cleaning or disinfecting solu-
tions.
►
Ensure that no fluids will penetrate the product.
CAUTION
Damage to, or destruction of the product caused by mechanical clean-
ing/disinfection!
►
Clean the product with wipe disinfection only.
►
Do not sterilize the product under any circumstances.
CAUTION
Damage to the product due to inappropriate cleaning/disinfecting
agents!
►
Only use cleaning/disinfecting agents approved for surface clean-
ing. Follow the manufacturer’s instructions for the respective
cleaning/disinfecting agent.
►
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids.
7.5
Wipe disinfection for electrical equipment
RT: Room
temperature
*
Recommended: Meliseptol® wipes sensitive (B. Braun)
Phase I
►
Remove any visible residues with a disposable disinfectant wipe.
Phase II
►
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
►
Observe contact time requirements (at least 1 min).
7.6
Inspection, maintenance and checks
►
Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
►
Set aside the product if it is damaged.
8.
Maintenance
To ensure reliable operation, the product must be maintained at least once
a year.
Contact your national B. Braun/Aesculap representative if you need these
services, see Chapter 10.
Phase
Step
T
[°C/°F]
t
[min]
Conc.
[%]
Water quality
Chemicals
I
Cleaning
RT
1
-
-
17 % Propan-1-ol,
0.23 % Didecyldimethylammonium chloride*
II
Wipe disinfection
RT
≥
1
-
-
17 % Propan-1-ol,
0.23 % Didecyldimethylammonium chloride*
Document No.: IFU - Version: 3.0 - Document ID: SOP-AIC-5001568 Date/Time Printed/Viewed: 2022-04-13 15:59 (CET)