Page | 72
Aeroqual AQM60 User Guide
* DATA - 4 bytes floating point data value. If the SENSOR_CODE is gas sensor, the value is concentration
in ppm. If the value is 9999, this means the corresponding sensor didn't give reading. Please refer to
Section 2 for detail.
* SENSOR_STATUS - 1 byte sensor status indication, refer section 3 for details.
* RB_TIMESTAMP - 6 bytes timestamp, in the format of ss:mm:hh DD-MM-YY, each field is 1 byte, or all
zero with no timestamp for SM50 basic controller.
* CHECKSUM - 1 byte the data stream's check sum - that makes the command stream total bytes sum be
zero.
3. AQM Monitor property configuration command, 4 bytes, followed by 26 bytes configuration data stream
(since V6.0):
RECEIVER, AQM_ID, SET_CONFIGURATION, CHECKSUM
* SET_CONFIGURATION - 1 byte command to configure AQM controller.
RECEIVER, AQM_ID, SET_CONFIGURATION, SENSOR_NUM, SENSOR_1, SENSOR_2, SENSOR_3,
SENSOR_4, SENSOR_5, SENSOR_6,
SENSOR_7, SENSOR_8, SENSOR_9, SENSOR_10, SENSOR_11, SENSOR_12,
SENSOR_13, SENSOR_14,
SENSOR_15, SENSOR_16, SENSOR_17, SENSOR_18, SENSOR_19, SENSOR_20,
AQM_STATUS, CHECKSUM
* AQM_ID - 1 byte value, indicate the AQM ID number, which range is 1 - 255.
* SENSOR_NUM - 1 byte, specify the valid numbers of sensor has been used by the AQM box.
* SENSOR_1, SENSOR_2, SENSOR_3, SENSOR_4, SENSOR_5, SENSOR_6, SENSOR_7,
SENSOR_8,
SENSOR_9, SENSOR_10, SENSOR_11, SENSOR_12, SENSOR_13, SENSOR_14,
SENSOR_15, SENSOR_16, SENSOR_17, SENSOR_18, SENSOR_19, SENSOR_20
- Specifies the sensor type has been used for the AQM box. The value will be any of Sensor type value.
* AQM_STATUS - 1 byte, used specify AQM auto report data and auto zero calibration, see section 2 for
details.
Example:
0x55 01 19 91
Followed by:
0x55 01 05 30 40 50 65 B0 00 00 00 00 00 00 00 00 00 03 CD
3A. Configuring AQM property acknowledgement command, 4 bytes:
MONITOR, AQM_ID, SET_CONFIGURATION, CHECKSUM
4. Request AQM Monitor property configuration command, 4 bytes,
RECEIVER, AQM_ID, GET_CONFIGURATION, CHECKSUM
* GET_CONFIGURATION - 1 byte configuration request command, its value please see section 2.
Example:
0x55 0x01 0x08 0xA2
5. Request AQM Monitor configuration command reply, 25 bytes (since V6.0):
MONITOR, AQM_ID, SENSOR_NUM, SENSOR_1, SENSOR_2, SENSOR_3, SENSOR_4, SENSOR_5,
SENSOR_6,
SENSOR_7, SENSOR_8, SENSOR_9, SENSOR_10, SENSOR_11, SENSOR_12,
SENSOR_13, SENSOR_14,
SENSOR_15, SENSOR_16, SENSOR_17, SENSOR_18, SENSOR_19,
SENSOR_20,AQM_STATUS, CHECKSUM
* AQM_ID - 1 byte value, indicate the AQM ID number, which range is 1 - 255.