7.3.4. Replacing Gas Treatment Media
The scrubber media can be changed by removing the front panel from the GTM. This can be achieved
without removing the entire GTM from the AQM. To release the front panel the two screws holding the
scrubber must first be removed. Pull the front cover out to expose the luer gas connection inside the GTM.
Once this gas connection has been released the media can be removed from GTM.
Gently lift up the large activated carbon scrubber to expose the two screws which are holding the media on
to the GTM front panel.
Disconnect the gas connection from the wall of the GTM front panel. Once this has been done then the
media can be removed and replaced with a new media kit. To install the new kit, simply reverse this
process.
Note: Remember to reconnect the luer connection before inserting the front panel back into the GTM