Page | 40
Aeroqual AQM60 User Guide
During a scheduled calibration or during a manual calibration, the status bar will indicate what the AirCal is
doing. If zero air is being applied the status bar will indicate Sample Port 0 (P0). The flow rate through the
MFM will be logged to the S.D. card and be displayed in the real time table. The inlet column will read Zero
Air.
When the AirCal is applying a span point the status bar will show Sample Px where x is the port number and
DR is the user defined dilution ratio. Both the flow through the MFM and the MFC will logged to the S.D. card
and be displayed in real time. The inlet column will display Px.