Page | 48
Aeroqual AQM60 User Guide
7.2.2. Particle Monitor
Procedure
Section
Frequency
Sample Flow Check
7.4.1.
Monthly
Purge Flow Check
7.4.2.
Monthly
Sheath Flow Check
7.4.3.
Monthly
Manual Zero Air Check
7.4.5.
Monthly
Fibre Span Check
7.4.6.
Monthly
Laser current check
7.4.7.
Monthly
Filter Changes
7.4.8.
6 to 12 months
Cyclone and Inlet Cleaning
7.4.9.
3 Months
Cyclone Disassembly
7.4.9.
12 Months
Optical sensor factory
calibration
Contact Aeroqual
24 Months
7.2.3. Profiler
Procedure
Section
Frequency
Sample Flow Check
7.5.1.
Monthly
Sheath Flow Check
7.5.2.
Monthly
Manual Zero Air Check
7.4.5.
Monthly
Filter Changes
7.5.4.
6 to 12 months
Inlet Cleaning
7.5.5.
6 months
Factory Calibration
Contact Aeroqual
Annual
7.3. AQM Maintenance Procedures
7.3.1. Replacing the Inlet Filter
The 5µm inlet filter should be replaced every 1-2 weeks depending on the specific conditions of the
monitoring environment. Unscrew and unplug the filter and replace. Ensure there are no leaks at the seal
when new filter is installed.