2.2.2. Connect Inlet Sub Assembly
Once the AQM60 has been connected the inlet sub assembly needs to be
installed to prevent unwanted materials, such as water and dust, being drawn into
the instrument. The sub assembly pipe is connected to the inlet at the top of the
AQM60 using a Kynar gland. The pipe needs to be securely inserted into the
Kynar gland and then the fitting needs to be tightened.
Ensure the two piece ferrule set is present inside the Kynar gland before screwing
it on, and be sure the orientation of the two pieces is correct as shown in the
image below. Failure to do this could result in a leak or could lead to water
entering the sample tubing which will permanently damage the sensor modules.
Kynar Gland
Neutral
(Blue)
Ground
(Green/Yellow)
Live
(Brown)
Retaining
clip
Retaining
clip