Aerogen
®
USB Controller System Instruction Manual
31
Warranty
The Aerogen warranty covers defects in manufacturing as follows:
• Aerogen USB Controller and Aerogen USB Controller AC/DC Adapter:
1 year
• Aerogen Pro: 1 year
• Aerogen Solo: Intermittent use for a maximum of 28 days and
continuous use for a maximum of 7 days.
All warranties are based on typical usage specified. There is no service life
associated with the Aerogen USB Controller System.
Life Of Products
The Aerogen USB Controller has been qualified for use for:
• 1460 doses (4 treatments per day, over a 1 year period.)
The Aerogen Pro nebulizer and components have been qualified for:
• 730 doses (4 doses per day.)
• 26 autoclave cycles (1 autoclave cycle per week, where the device is
assumed to be in service for 50% of the time over a 1 year period.)
• 104 disinfecting treatments (2 per week, over a 1 year period.)
The Aerogen Solo nebulizer has been qualified for:
• Intermittent use for a maximum of 28 days (4 treatments per day.)
• For continuous use the life of the Aerogen Solo nebulizer and the
Continuous Nebulization Tube Set have been validated for use for a
maximum of 7 days.
The user should note that use of the Aerogen technology in excess of
these periods is not qualified by Aerogen.