23
Functional Test
Perform a functional test prior to first use, or after each sterilization
(Aerogen Pro only) to verify correct operation. This test is to be carried out
prior to inserting the nebulizer into a circuit or accessory.
1. Visually inspect each part of the system for cracks or damage and
replace if any defects are visible.
2. Pour 1-6 mL of normal saline (0.9%) into the nebulizer.
3. Connect the nebulizer to the Aerogen USB Controller. Connect the
Aerogen USB Controller to the Aerogen USB Controller AC/DC Adapter.
4. Press and release the On/Off power button and verify that the 30
Minute Mode indicator light is green and that aerosol is visible.
5. Disconnect the nebulizer from the controller. Verify that the amber Error
Indicator lights. Reconnect the nebulizer to the controller.
6. If using the Aerogen Solo press the On/Off power button again to turn
the system off. Press and hold the button for at least 3 seconds. Verify
that the 6 Hour Mode indicator light is green and that aerosol is visible.
7. As with step 5 above, disconnect the nebulizer from the controller.
Verify that the amber Error Indicator lights. Reconnect the nebulizer
to the controller.
8. Turn the system off and verify that the 30 Minute Mode and 6 Hour
Mode indicator lights are off.
9. Discard any remaining liquid before patient use.