Aerogen
®
USB Controller System Instruction Manual
51
Accessories
T-Pieces - Connection To A Breathing Circuit
Adult & Pediatric Circuit
For adult and pediatric patients, connect
the nebulizer with T-piece into the
inspiratory limb of the breathing circuit
before the patient Y.
Alternative Pediatric Circuit
Connect the nebulizer to 10mm pediatric
breathing circuits with the 15mm pediatric
T-piece and the pediatric adapters. This
can be positioned approximately 30 cm
(12 in.) back from the patient Y.
Dry Side of the Humidifier
The Aerogen Solo can be placed on
the dry side of the humidifier as shown.
The Aerogen Solo can be used with a
nasal interface in this configuration. The
Aerogen Pro is not recommended for
use on the dry side of the humidifier.
Note:
Always perform a leak test of the breathing circuit after inserting or
removing the nebulizer.
Follow ventilator manufacturer instructions for performing a leak test.
For additional T-piece Adapters visit www.aerogen.com for full
parts list.
Содержание Aerogen USB Controller
Страница 2: ......
Страница 40: ...38 Aerogen Aerogen USB Controller System Instruction Manual for use with Aerogen Solo and Aerogen Pro HOME USE...
Страница 70: ...68 Aerogen This page has been intentionally left blank...
Страница 71: ......