Guidance and manufacturer’s declaration – electromagnetic immunity for the
Aerogen USB Controller System that is not life supporting
The Aerogen USB Controller System is intended for use in the electromagnetic
environment specified below. The customer or the user of the Aerogen USB Controller
System should assure that it is used in such an environment.
Immunity Test
IEC/EN 60601
Test Level
Compliance
Level
Electromagnetic
Environment - Guidance
Electrostatic
discharge (ESD)
IEC/EN
61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be at
least 30%.
Electrical fast
Transient/burst
IEC/EN
61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge
IEC/EN
61000-4-5
±1 kV line(s) to
line(s)
±2 kV Line(s) to
earth
±1 kV line(s) to
line(s)
±2 kV Line(s) to
earth
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC/EN
61000-4-11
<5 % Ut
(>95 % dip in Ut)
for 0.5 cycle
40 % Ut
(60 % dip in Ut)
for 5 cycles
70 % Ut
(30 % dip in Ut)
for 25 cycles
<5 % Ut
(>95 % dip in Ut)
for 5 sec
<5 % Ut
(>95 % dip in Ut)
for 0.5 cycle
40 % Ut
(60 % dip in Ut)
for 5 cycles
70 % Ut
(30 % dip in Ut)
for 25 cycles
<5 % Ut
(>95 % dip in Ut)
for 5 sec
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the Aerogen USB Controller
System requires continued
operation during power
mains interruption, it is
recommended that the
Aerogen USB Controller
System be powered from an
uninterruptible power supply.
Power
frequency
(50/60 Hz)
Magnetic field
IEC/EN
61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Note
: Ut is the A.C. mains voltage prior to application of the test level.