54
Aerogen
®
Cleaning of the Aerogen Pro Nebulizer
The Aerogen Pro components are:
• Nebulizer (including filler cap)
• T-piece (including T-piece plug) for adult and pediatric
Manual Cleaning of Aerogen Pro & Accessories
1. Ensure there is no medication remaining in the device.
2. Remove nebulizer from T-piece. Remove filler cap from nebulizer.
3. Clean all parts with warm water and mild liquid detergent.
4. Rinse parts with sterile water.
5. Shake excess water from parts and allow parts to fully air dry.
Cleaning is recommended to be carried out between treatments on the
same patient.
Warning:
Do not use abrasive or sharp tools to clean the nebulizer unit.
Boiling the Aerogen Pro & Accessories
1. Rinse the Aerogen Pro nebulizer components under running hot tap
water after every use.
2. Shake off excess water, and allow parts to fully air dry on a clean, dry
towel, out of the reach of children.
3. Bring a saucepan of DISTILLED water to the boil and allow it to continue
to boil.
Note:
Using regular tap water to boil the nebulizer will greatly reduce the
life of the nebulizer.
4. Carefully immerse the Aerogen Pro nebulizer in the boiling water.
Leave the nebulizer immersed in the boiling water for a maximum of
20 minutes.
Содержание Aerogen USB Controller
Страница 2: ......
Страница 40: ...38 Aerogen Aerogen USB Controller System Instruction Manual for use with Aerogen Solo and Aerogen Pro HOME USE...
Страница 70: ...68 Aerogen This page has been intentionally left blank...
Страница 71: ......