Table F‑2. Converter box and analog box UPP commands (Continued)
Parameter
Command
Description
The converter box responds with a 2-digit value
from
00
to
08
identifying the number of sensor
heads connected.
Display
orientation
<AA>ro<VAL>
Sets the orientation of the converter box built-in
display.
Valid values are 0 to 3.
0 = 0°
1 = 270°
2 = 180°
3 = 90°
Automatic
addressing
<AA>AD<VAL>
Automatically sets the sensor head address for
all connected sensor heads.
Valid values are 0 to 2.
0 = Disables automatic addressing (the
converter box responds with
no
)
1 = Sets the address of all connected sensor
heads in the order shown on the converter
box display or in the Infra600 software (A1
to A8)
2 = Resets all sensor head addresses
Automatic
addressing
option
<AA>AA<VAL>
Toggles the automatic addressing function of the
converter box.
Valid values are 0 and 1.
0 = Disable automatic addressing
1 = Enable automatic addressing
Repeat request
<AA>rp<VAL>
Defines how often a request to a sensor head
is repeated (in case of no response or a wrong
answer) before an error message is shown
(timeout).
Valid values are 00 to 99.
Restart
converter box
<AA>re
Restarts the converter box (does not affect any
settings).
Read
PROFINET
address
<AA>ip
Reads the address of the PROFINET module
(8-digit hexadecimal ACSII).