Table F‑2. Converter box and analog box UPP commands (Continued)
Parameter
Command
Description
Wait time (t
w
)
<AA>tw<VAL>
Sets the wait time for the converter box. The
set value represents a fraction in relation to the
selected baud rate.
Valid values are 00 to 99.
Example: 02 (at a baud rate of 9600) = 2/9600 s
(0.00021 s)
Interface
<AA>in<VAL>
Sets the interface for the converter box.
Valid values are 1 and 2.
1 = RS-232
2 = RS-485
Temperature
scale
<AA>fh<VAL>
Sets the temperature scale to Celsius or
Fahrenheit for all the converter box parameters.
Valid values are 0 and 1.
0 = Temperature in °C
1 = Temperature in °F
Model and
firmware
<AA>ve
Reports the model number and firmware version
for the converter box.
The pyrometer responds with a 6-digit value:
VVMMYY
•
VV
= Device ID
45
= Series 600 converter box
•
MM
= Month that the firmware was released
•
YY
= Year that the firmware was released
Serial number
<AA>sn
Reports the unique serial number for the
converter box.
The pyrometer responds with a 7-digit value.
Reference
number
<AA>bn
Reports the reference number for the converter
box.
The pyrometer responds with a 6-character
hexadecimal value (equivalent to the 7-digit
reference number).
Keys lock
<AA>lk<VAL>
“l” is alpha.
Sets the lock status of the converter box keys for
the built-in display.